Event logging – YSI 5500D-1 User Manual

Page 57

Advertising
background image

YSI 5X00

113

YSI 5X00

112

In the above sensor log screen shot Aux 3 is assigned as temperature for DO 1. DO

4 and Aux 1, 2, 4, 5, & 6 are disabled.

Event Logging

Menus → System → Data Logging → Event Logging

Control, Alarm, and Timer relays assigned to 5X00 systems create events when they

turn on and off. In addition, AC power failures, ethernet device error messages,

and slave network failures create events. These events are stored in the Event Log.

Event logs identify the 5X00 system, are time and date stamped, and record relay

action (“on” or “off”).

1. Enter Event Logging menu

Menus → System → Data Logging → Event Logging

2.

Use

to scroll and highlight submenus.

3.

Press

to select.

4. Press

to turn off or on.

5. Press

to save setting and return to Run Screen.

Press

to view 25 most recent event logs.

Press

to view active (“on”) non timer events. See Control Menu - page 170

for additional information.

Notes:

- The General Alarm does not create an event whether a general alarm relay is

enabled or not.

- Events are not recorded when relays de-energize due to system-wide events.

System-wide events include calibration, Svr Hold, factory resets, power cycle,

30 second sensor system reset, relay test, and disabling systems.

Configuring the 5X00

- Event logs record the default system name not user configured system

name.

- Event logs do not record relay number or user configured relay name.

- Aux digital events will never display “alarm” regardless if alarms are enabled,

see below screen shot on right.

Clear Logs

Menus → System → Data Logging → Clear Logs

Use the Clear Log option to permanently delete all event and/or sensor log records

from the 5X00.

Before permanently deleting sensor and event logs, poll all data using AquaManager.

AquaManager will store all logs in PC database.

1. Enter Clear Log menu

Menus → System → Data Logging → Clear Logs

2.

Use

and

to scroll submenu.

3.

Press

to select.

4. The instrument will prompt, “Are You Sure?” Press one of the following

softkeys to confirm:

.

5. Selecting “Yes” deletes logs after approximately three seconds.

6. Selecting “No” or “Cancel” does not delete logs and returns to the Data

Logging menu.

7. Press

to save setting and return to Run Screen.

Notes:

- Logs cannot be deleted from the 5X00 instrument using AquaManager.

- Calibration and Configuration

logs cannot be cleared from 5X00.

- Clear data logs are updated immediately to the system and not when

is pressed. YSI recommends pressing

in this instance in order to save

any other changes that may have been made to the configuration. Exception -

Configuring the 5X00

Advertising
This manual is related to the following products: