Data Logging

Menus → System →Data Logging
The 5X00 saves information in calibration, sensor, and event logs. Press

at the Run Screen to view the 25 most recent records for each log. The 5X00 sensor

log stores 2000 records. The calibration log can store 500 records. And the event

log stores 1250 records. Once a 5X00 log is full, it will begin overwriting the oldest

records with new records (data wrapping).

Notes:

- Use AquaManager to store and analyze 5X00 logs. To avoid gaps in data, use

polling intervals based on the user defined data logging configuration. Set

the data log interval so that analyzing the data is meaningful.

- AquaManager users can view Configuration events in the Configuration log.

Configuration

events are created when certain 5X00 configuration changes are

made. The configuration log is not accessible at the 5X00. See AquaManager

Online Help for additional information.

Data Logging

Menus → System →Data Logging→Data Logging

The 5X00 sensor log stores records for sensors and aux systems. Sensor logs are time

and date stamped based on the user configured logging interval (1-255 seconds/

minutes/hours).

Configuring the 5X00

1. Enter Data Logging menu

Menus → System →Data Logging→Data Logging

2.

Use

to scroll and highlight submenus.

3.

Press

to select.

4. Enable and configure submenus.
5. Press

to save setting and return to Run Screen.

Press

to view the 25 most recent sensor logs

Notes:

- Aux digital “closed” is logged as Aux 0.00. Aux digital “open” is logged as

Aux 1.00.

- The 5X00 sensor log records 0.00 resolution for aux analog inputs regardless

of the selected display format.

- DO values are stored in mg/l & %Sat regardless of selected display format.

- Sensor logs record the default system name and not the user configured

system name.

- Aux temperature values are not accessible at the 5X00. Temperature values

can only be obtained using AquaManager. Values will be shown in the Aqua-

Manager data log in °C regardless of the selected display format.

- Temperature values are only available and stored for temperature systems as-

signed to a DO system. These values are not available at the 5400. The values

can only be viewed with AquaManager.

- Temperature values are available for either a ODO temperature sensor or a

stand alone aux temperature systems. These values are not available at the

5500. The values can only be viewed with AquaManager.

- Changes to the data logging interval take effect after the expiration of the

interval being changed. Example - if the interval is changed from one hour to

one minute, one minute logging will begin after the expiration of the current

one hour interval. Power cycle the 5X00 to make a change to the data logging

interval occur immediately.

- Menu configurations are saved to the system by pressing

to return

to the Run Screen and not when the change is made in the menu. Excep-

tion - Date and time and clear data logs are updated immediately to the
system, not when

is pressed.

- The sensor log contains values for all sensor and aux systems regardless if the

system is disabled.

- 5X00 sensor log values for disabled systems and systems without sensors are

not valid.