Wire dc power and backup power, Directions to wire dc power/dc backup power – YSI 5500D-1 User Manual

Page 25

Advertising
background image

YSI 5X00

49

YSI 5X00

48

Wire DC Power and Backup Power

The 5X00 may be powered by most regulated 9.0-16.5 VDC source that provides

800mA of current and is isolated from mains supply by double or reinforced insula-

tion. The DC power source is user supplied and could be lead acid, gel cell, or UPS

(with 12VDC transformer) external batteries. DC power wires are terminated at

location B on the I/O Board - Figure 3.18 (Common lettering references are used

for both 5400 and 5500 models).

Notes:

- The 5X00 does not charge batteries. Quality assurance maintenance proce-

dures should be established if batteries are to be fully powered when used as

back-up power source.

- Power supply voltage above 16.5 VDC may permanently damage the 5X00.

- When back up voltage falls below 9.0 volts, the 5X00 ceases to operate properly.

- Some switching DC power supplies not supplied by YSI may result in noisy

readings.

Directions to Wire DC Power/DC Backup Power

WARNING: Disconnect external power to the unit before wiring.

Follow all manu-

facturer’s safety and installation instructions when providing power to the 5X00

via a DC power supply.

AVERTISSEMENT  : déconnectez l’alimentation externe de l’unité avant

d’effectuer le câblage. Suivez toutes les consignes de sécurité et d’installation du

fabricant lorsque vous fournissez de l’alimentation au modèle 5X00 par un bloc

d’alimentation à courant continu.

1. Perform steps 1-3 of 5X00 installation.

2. Open front panel - page 42.

3. Feed DC power cable through drilled hole in rubber grommet to location

B. Terminal strip is removable.

CAUTION

: Run high and low voltage cables through separate bulkhead and

conduit.

4. Terminate DC power wires to pins (-) and (+) according to manufacturer

instructions and any applicable local electrical codes.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5400 is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.18

5. Complete other wiring to 5X00.

6. Close front panel - page 43.

7. Test DC power.

Test DC Power

Connect DC power cable to power source. The enabled system values will display

on Run Screen after initial YSI splash screen. Verify DC power at the Menus →

System →Version menu - page 135.

See Power Fail Menu - page 128 for DC low battery trigger monitor and alarm

information.

Note:

- The 5X00 will not power up if DC power supply wires are terminated incor-

rectly.

Installation and Wiring

Installation and Wiring

Advertising
This manual is related to the following products: