Accessing the i/o board, Open front panel to access i/o board, Close front panel – YSI 5500D-1 User Manual

Page 22

Advertising
background image

YSI 5X00

43

YSI 5X00

42

Accessing the I/O Board

The I/O board is accessible when the 5X00 front panel is removed. The letters on

the metal plate, which is attached to the lower board assembly, identify ground wire

termination locations and pluggable terminal strip locations on the I/O board. Letters

on the metal plate correspond to the wiring diagram located on the inside of the top

panel - figure 3.14. A ribbon cable connects the display board (mounted in the top

panel) to the I/O board. When the 5X00 top panel is removed, a lanyard protects

against stress on the ribbon cable. See Open Top Panel to Access I/O Board - page 42.

Figure 3.14

Open Front Panel to Access I/O Board

WARNING: Disconnect external power to the unit before doing any wiring.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer

un câblage quelconque.

1. Loosen the 4 mounting screws connecting the front panel to back enclosure.

Captive screws remain in front panel after loosening.

2. Carefully lift the 5X00 front panel from the back enclosure. The front

panel should be placed to the left of enclosure. Allow the lanyard to hold

the weight of the front panel. The rubber gasket seated in the front panel

should remain in the front panel lid.

Close Front Panel

1. Verify all installed components and peripheral equipment have secure

connections and that there are no bare wires that could cause a short inside

the enclosure.

2. Tighten bulkhead fittings after connecting all 5X00 wiring. (Hand tighten

fittings around rubber grommet. Tighten bulkhead fittings into 5X00 en-

closure with wrench being careful to not overtighten.)

3. Follow information in Sealants, Desiccants and Securing the Monitor Section

- page 71 to safeguard against damage to internal electronic components.

4. Make certain the gasket in front panel lid is in place and not twisted or

damaged.

5. Replace front panel making sure that ribbon cable is not trapped in the gasket

channel before inserting mounting screws.

6. Tighten mounting screws, making sure not to cross thread. The screws are

stainless steel, and the receiving threads are brass. Do not over-tighten!

WARNING: The 5X00 utilizes sensitive solid-state devices that can be damaged

by static shock. Installers must observe accepted ESD (Electro-Static Discharge)

procedures while connecting cabling to the 5X00 I/O plate or to any other internal

component or damage may result.

AVERTISSEMENT : le modèle 5X00 utilise des appareils transistorisés sensibles

pouvant être endommagés par des décharges d’électricité statique. Les installateurs

doivent observer les procédures de protection acceptées contre les décharges

électrostatiques lors du câblage à la carte d’E/S du modèle 5X00 ou à tout autre

composant interne afin d’éviter tout risque de dommage.

Installation and Wiring

Installation and Wiring

Advertising