Email alarming, Rs232 communication – YSI 5500D-1 User Manual

Page 35

Advertising
background image

YSI 5X00

69

YSI 5X00

68

Notes:

- Firmware can only be uploaded to an instrument using a local RS232 connec-

tion. See Downloader - page 119.

- On a network, install and configure RS232 or the serial to ethernet device

hardware on the master node only.

Email Alarming

Email alarming can be configured in 1 of 2 ways:

from PC running AutoPoll Program that is included with AquaManager

software (PC must have internet connection). This is referred to as Aqua-

Manager email.

via a serial to ethernet* device with TCP/IP connection at the master 5X00.

This is referred to as 5X00 email.

*Serial to ethernet connection requires access to an SMTP server accessible by the

local network.

RS232 Communication

Use Category 5e patch cable from RS232 port at location J - figure 3.31 of I/O

Board to PC using RJ45 to DB9 adapter. (Common lettering references are used

for both 5400 and 5500 models).

Directions to Wire Direct RS232 PC Connection

WARNING: Disconnect external power to the unit before wiring.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer

un câblage quelconque.

1. Perform steps 1-4 of 5X00 installation and steps 5 - 8 as required for

the application.

2. Open front panel - page 42.

3. Feed direct connect cable through drilled hole of rubber grommet and

to location J.

CAUTION

: Run high and low voltage cables through separate bulkhead and

conduit.

4. Connect one end of communication cable to RJ45 at location J. Connect

other end to RS232 PC com port using RJ45 to DB9 adapter figure 3.31.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5400 is po

w

er

ed

. S

er

ious damage can oc

cur

.

figure 3.31

5. Complete other wiring to configure 5X00 system.

6. Close front panel - page 43.

7. Test RS232 communication.

Notes:

- When using a USB-Serial Adapter, verify that the port number assigned dur-

ing driver installation is the same port configured when using AquaManager

and/or the Firmware Updater

- Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS YSI

p/n

773

(64-bit).

- When making RS232 cable runs greater than 100 feet (30.5 m), use an RS232

Line Extender/Booster. Mount the booster in an area where it is protected

from moisture.

Test RS232 Communications

Apply power to 5X00. Insert node into AquaManager program. See AquaManager

Help to insert node and configure communication in PC application. Poll node. If

wired and configured properly, AquaManager will obtain 5X00 configuration and

5X00 logs.

Installation and Wiring

Installation and Wiring

Advertising
This manual is related to the following products: