YSI 5500D-1 User Manual

Page 48

Advertising
background image

YSI 5X00

95

YSI 5X00

94

1. Install serial to ethernet device - page 65.
2. The Ethernet icon displays at the Run Screen when the serial to

ethernet device is properly installed.

3. Enter Ethernet menu

Menus → System → Communications → Ethernet

4.

Use

and

to scroll and highlight submenus.

5.

Press

to select submenu.

6. Configure submenus as needed.
7. Press

to save configuration and return to Run Screen.

Notes:

- The CC/Email address must be configured when using 5X00 email alarming.

Specific alarm system information will not be sent if the CC/Email address

menu is not configured with a valid email address.

- If Autopoller is being used for AquaManager email alarming on a TCP/IP

connection, it is strongly recommended that the 5X00 email alarming using

the serial to ethernet device be disabled. When the ethernet device sends

email alarms, all Autopoller SMS communication is suspended. This could

potentially delay any AquaManager alarm emails from being sent.

- To disable the serial to ethernet device from sending alarms, clear all four

email addresses from the Communication → Ethernet menus. Clear the email

addresses by using the “Spc” (space) key on the numeric keypad. Save changes
at numeric keypad and press

to save configuration and return to Run

Screen.

Configuring the 5X00

DHCP

Menus → System → Communications → Ethernet → DHCP

Enable DHCP if you are using the serial to ethernet device only for 5X00 email

alarming and not for AquaManager remote TCP/IP access.

Notes:

- When DHCP is enabled, the 5X00 IP address is not updated at the IP address

menu.

- With DHCP enabled, the serial to ethernet device acquires SMTP and DNS

addresses during initialization.

- AquaManager requires a static IP address; therefore, DHCP must be disabled

when using AquaManager. Consult the network administrator to obtain a

static IP address.

- Power cycle the 5X00 if DHCP configuration changes from disabled to en-

abled..

IP Address

Menus → System → Communications → Ethernet → IP Address

The IP address can only be configured when DHCP is disabled.

Notes:

- AquaManager users should program the IP Address either by force serial port

using AquaManager or locally at the 5X00. Poll node using RS232 commu-

nication . Next upload configuration information including: static IP, port,

subnet, and gateway addresses.

- When DHCP is enabled, the IP address is not updated at the IP address

menu.

- Power cycle the 5X00 whenever a new IP address is configured.

- Numbers between 0-255 are valid for each of the four (4) IP Address fields.

Configuring the 5X00