Odo cal k, 5500d-01, 5500d-02 and 5500d-04) – YSI 5500D-1 User Manual

Page 59

Advertising
background image

YSI 5X00

117

YSI 5X00

116

1. Enter Display Settings menu

Menus → System → Display Settings →Display Format

2.

Use

and

to scroll and highlight submenus.

3.

Press

to select.

4.

to save configuration and return to Run Screen

.

Notes:

- At 1x2 & 2x2 display settings, analog aux values over 7 digits will overlap on to

the adjoining sensor window. Adjust aux display format to lower resolution.

- Because of font limitations, 1X2 and 2X3 display formats do not show all

20 characters of a 20 character name. Typically only 13 characters are dis-

played.

ODO Cal K

(5500D-01, 5500D-02 and 5500D-04)

Menus → System → ODO Cal K →DO1-DO4

The ODO Cal K system menu appears only in 5500D-01, 5500D-02 and 5500D-04

instruments. This menu is for viewing and uploading DO sensor cap calibration

codes. Uploading ODO sensor cap calibration codes is required whenever a sensor

cap is replaced.

Cap calibration codes are not stored in the 5500D instrument but rather are stored

5500

Configuring the 5X00

in the ODO sensor. Upon selecting DO-1, DO-2, DO3 or DO-4 at the ODO Cal

K menu communication initiates between the instrument and the sensor. An ODO

sensor has six cap calibration codes. They are known as K1, K2, K3, K4, K5 and KC.

Direction to view or upload ODO sensor cap calibration codes:

1. Enter ODO Cal K menu

Menus → System → ODO Cal K →DO-1, DO-2, DO-3, DO-4

2. Select DO -1, DO-2, DO-3 or DO-4

ODO sensor cap calibration codes are displayed

If the 5500D is unable to obtain the cap calibration codes from the ODO sensor the

below error message is displayed. Typically a retry after several seconds will display

the ODO sensor cap calibration codes. If the calibration codes cannot be displayed

confirm that the ODO sensor is displaying a value at the run screen.

3. To upload sensor cap calibration codes select K1, K2, K3, K4, K5 or KC.

4. Use alpha/numeric keyboard to enter the values exactly as provided on

the sensor cap instruction sheet sent from YSI.

Configuring the 5X00