PID/PWM Alarm System

Alarm values set the acceptable control range. Enable and configure system(s) alarm

to activate alarm output devices(s), view alarm icons, hear audible alarm and/or

receive alarm notifications sent via email or cell phone (SMS) - page 91. Config-

ure high and low alarm values to set alarm thresholds. Alarm system(s) becomes

active when value is ≤ low alarm value or ≥ high alarm value. Enabled alarm relays

energize if alarm condition exists for the general alarm hold off time. Alarm relays
are energized until one of the four softkeys

is pressed

regardless of sensor value. Pressing one of four softkeys resets the alarm system.

Alarm systems become active again if, after the alarm hold off period, the alarm

condition still exists. Alarm hold off is configured in the General Alarm menu.

When the General Alarm is enabled, (with or without enabled relay), the general

alarm system becomes active when any system alarm is active. See General Alarm

- page 124.

Notes:

- It is important to keep the 5X00 at the Run Screen during normal opera-

tion. Alarm functionality (email alarms, activation of enabled alarm relays,

alarm icons, and buzzer) is suspended when 5X00 is not at Run Screen.

- Alarm icon(s) display at the Run Screen when alarm condition(s) exists for

longer than the general alarm hold off time. See Icons - page 78.

- Press

at Run Screen to display active sensor and aux system relays.

-

to display events created by control and alarm relays - see Event Log

- page 112.

- Alarm relays are not energized if the 5X00 DO sensor value reads “Over” or

“Under.”

- “Over” and “Under” are not displayed for aux temperature systems when

temperature values are outside the temperature range specification - see tem-

perature range specification - page 15. Temperature system alarms will not

operate correctly when values are outside the temperature range specification.

- See Wire Relays - page 56 for information on wiring alarm output devices.

30 Second Aux and Sensor System Hold Off

Exiting aux and sensor menus after making certain configuration changes places the

system in a 30 second hold. During the 30 second sensor hold, control and alarm

systems are not active. System(s) becomes active, if, after 30 seconds, the values

are reported in control or alarm range. This 30 second hold off allows sensors and

aux devices time to stabilize before control and/or alarm systems become active.

Example - DO sensor menu is entered and the control high set point value is changed
while a control relay is energized. When

is pressed to return to the Run

Screen, the DO control relay de-energizes regardless of sensor value. The control

relay re-energizes if, after the 30 second hold, the value is still in control range.

Reconfiguration of the following does not place the system in 30 second hold off:

- DO systems - display format, probe type, membrane type, ppt, elevation, temp

source, and temp display format;

- Temperature system - display format, chiller on time, chiller on time units,

chiller hold off time, chiller hold off time units , set points, enable/disable

control relays.

Notes:

- Configuration changes place only the system that has been reconfigured on

30 second hold.

- No “off” event log is created when energized system relays de-energize due

to 30 second sensor hold off.

- Control icons and control messages continue to display during the 30 second

sensor system hold at the Run Screen regardless if relays are energized.

Aux Hold Off (non temperature aux systems)

Aux Hold Off (user configured) is the length of time the control or alarm condition

must exist before the Aux relay energizes.

Notes:

- To avoid control or alarm for spurious events , configure the Alarm Hold Off

> than 0.

- If General Alarm is enabled, it is triggered after the Aux hold off time and not

the general alarm hold off time.

Aux Digital

Configure digital control/alarm (open/closed), relay, alarms, and hold off in the Aux

1 - Aux 6 Sensor Setup Menu. Use digital input type for control, or alarm, based on

either a switch open or switch closed condition. Only one relay can be configured in

an aux digital system. Enabled relay will energize based on the condition configured

in the control/alarm on menu.