Wire temperature sensor(s), Directions to wire temperature sensor(s) – YSI 5500D-1 User Manual

Page 28

Advertising
background image

YSI 5X00

55

YSI 5X00

54

5. Wire temperature sensors. Temperature sensors are wired as auxiliary inputs.

See Temperature Sensor(s) - page 54.

6. Complete other wiring to configure 5X00 system.

7. Close Front panel -page 40.

8. Calibrate DO sensor - see Calibration “DO” on page 85.

CAUTION

: Do not ground the probe body.

Test DO/Temperature Sensors

Apply power to 5X00. See DO Sensor Setup - page 154 to configure DO sensor

inputs. See Temperature Setup - page 150. Calibrate the DO sensor, see DO

calibration - page 82. Verify changes in sensor values occur when the sensor is

placed in environments of different oxygen concentration and temperature. For

example, the DO reading in saturated air for calibration versus the DO reading

when the sensor is placed in a zero oxygen environment (mix 1 gram sodium sulfite

in 0.5 liter of water).

Wire Temperature Sensor(s)

The 5X00 system will support four Temperature/Aux Analog inputs. Temperature

sensors have 2 wires and are terminated at location (5500) L (5400) on I/O

Board - figure 3.20. Temperature sensor(s) must be thermistors with 10K at 25˚C,

see Appendix 6 Alpha A Curve - page 236.

Directions to Wire Temperature Sensor(s)

WARNING: Disconnect external power to the unit before wiring.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer

un câblage quelconque.

1. Mount and install 5X00 and probes - page 27 and page 33.

2. Open front panel - page 42.

3. Feed temperature probe cable through drilled hole in rubber grommet

to location (5500) L (5400). The terminal strip is removable.

CAUTION

: Run high and low voltage cables through separate bulkhead and

conduit.

4. Terminate temperature sensor wires to Aux 3, Aux 4, Aux 5, and/or Aux

6 system. There is no polarity for temperature sensor wires; therefore,

(+) and (-) terminals are interchangeable. Make sure the temperature

sensor wires are terminated to the (+) and (-) terminals for the Aux

system (3, 4, 5, 6) you are configuring.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5400 is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.20

5. Complete other wiring to configure 5X00 system.

6. Close front panel - page 43

7. Test temperature sensor.

Note:

- Temperature sensors wired to Aux 3, Aux 4, Aux 5, and Aux 6 are displayed

on the 5X00 by default as Temp 1, Temp 2, Temp 3, and Temp 4 respectively.

For instructions on changing the default display settings, see section Name

Devices/Sensors in Chapter 4 Configuring the 5X00.

Test Temperature Sensor

Apply power to 5X00. See Temperature Set up- page 150 to configure temperature

system. Place the temperature sensor in solutions of varying temperature and verify

the sensor readings with another temperature sensor.

Installation and Wiring

Installation and Wiring

Advertising
This manual is related to the following products: