Control Board

The control board is mounted underneath the I/O Board. The serial to ethernet

device and lithium battery are installed on the control board. Complete installation

instructions, including accessing control board, are provided with the optional serial

to ethernet device. Figure 3.15 shows the control board and I/O Board removed

from the 5X00 enclosure.

Figure 3.15

Step 3 Ground 5X00

WARNING

: Ground the 5X00 to avoid possible electrical shock or damage to

the equipment.

AVERTISSEMENT : mettez le modèle 5X00 à la terre afin d’éviter tout risque de

choc électrique ou d’endommagement de l’équipement.

Ground the 5X00 at location G on the metal plate. The metal plate is attached to the

lower board assembly. Use locations E and F to ground the DO sensor ground/drain

wires - figure 3.16. (Common lettering references are used for both 5400 and 5500

models). Titanium ground rods can also be used to ground peripheral equipment.

In addition to grounding the 5X00, all tanks should be electrically grounded via a

ground probe.

Directions to Ground 5X00

WARNING: Disconnect external power to the unit before wiring.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer

un câblage quelconque.

1. Perform steps 1 - 2 of 5X00 installation.

2. Open front panel - page 42.

3. Remove screw from location F or G.

4. Feed ground wire through drilled hole in rubber grommet and to location F

or G on metal plate covering I/O Board. Ground wire gauge must conform

to all locally applicable electrical codes.

CAUTION: Run high and low voltage cables through separate bulkhead and

conduit.

5. Install a ring (eyelet) or spade (fork) terminal onto the ground bare wire

before terminating ground wire to F or G.

6. Return screw at F or G over terminal connection of ground wire.