Notes:

- I/O board and control board (including DO daughter board) are attached to

a metal plate and are referred to as the “lower board assembly”. See Control

Board - page 44.

- No 5X00 components or peripheral devices are wired to the display or DO

daughter board.

Most components including: power, Aux input, network, relay output, and RS232

communication components are terminated on the I/O board. 5X00 ground wire

and sensor ground wires (also called drain wires) are terminated to the metal plate

that is attached to the lower board assembly. The optional serial to ethernet device is

installed on the control board. The front panel must be removed when wiring 5X00

components and peripheral devices. See Accessing the I/O Board - page 42. It is

recommended that some system configuration and testing be done in conjunction

with hardware installation. This will ensure that the 5X00 system is wired correctly.

Necessary configuration information (from Chapter 4 - “Configuration”) is referenced

in the applicable test system procedure.

WARNING: Follow all safety information and local electrical codes when wiring

5X00 system components and peripheral devices. Proper wire gauge should be

determined based on voltages and wire/cable length. Incorrect wiring can result

in damage to you or equipment. Ground loops can also result due to improper

wiring.

AVERTISSEMENT : suivez toutes les consignes de sécurité et les codes électriques

en vigueur lors du câblage des composants du système 5X00 et des appareils

périphériques. Le calibre correct des câbles doit être déterminé en fonction des

tensions et de la longueur des câbles/fils. Un câblage incorrect peut entraîner des

blessures et endommager l’équipement. Des boucles de mise à la terre peuvent

également être provoquées par un câblage incorrect.

5X00 Bulkhead Fittings

Five bulkhead fittings [5@ 3/8”(.9525cm), 2 @ 3/4” (1.905cm), and 3@ 1/2” (1.27cm)]

located at the bottom of the 5X00 enclosure provide watertight connections for

5X00 system component wiring. Rubber grommets must be “pushed out” of the

compression fitting housing and drilled for wiring installation. The drill bit diameter

should be slightly smaller than the wire diameter so that a watertight seal can be

made. Rubber grommets can be frozen (put in a freezer for 1 hour) and then easily

drilled to the required diameter. Some pre-drilled rubber grommets are supplied

with the 5X00.