Network troubleshooting, 5x00 systems troubleshooting, Firmware updater troubleshooting – YSI 5500D-1 User Manual

Page 102

Advertising
background image

YSI 5X00

203

YSI 5X00

202

Issue

possible cause

possible solution

Serial to ethernet device did

not initialize properly. Check

the event log for “unable to

initialize error”.

disconnect power to 5X00,

verify installation of serial

to ethernet device on lower

board and repower

faulty ethernet serial to eth-

ernet device

replace serial to ethernet

device

Are the alarms enabled in

sensor menu? Is the general

alarm enabled?

see Sensor Setup Alarms

Values - Set point control

mode - page 140 and PID/

PDW control mode - page

146.

see General Alarm - page

124..

no email alarm received

for slave alarm condition

Are slaves enabled on net-

work?

enable slaves at Menus →

System → Communications

→ Network →Master

no test email alarm mes-

sage received

exit to run screen from com-

munication menu not done

exit to run screen to process

test email request

Network Troubleshooting

Issue

possible cause

possible solution

no communication to

networked unit(s)

Are master/slave menus

properly configured?

See Networking - page

98..

Is termination relay en-

abled in 1st & last 5X00 on

network?

See Termination Relay -

page 101..

Is proper communication

method configured and

wired properly?

See Networking - page 98.

See Wire Communication

Equipment - page 67..

Are slaves enabled and recog-

nized at master?

enable slaves at Menus →

System → Communications

→ Network →Master

more than one device is

configured as a master

verify configuration at

Menus → System → Com-

munications → Network

→Master/Slave menus on

all nodes.

Troubleshooting

Email Alarm Troubleshooting continued

slow network perfor-

mance

nicked wires;

improper wiring

Verify network wiring -

See Wire Network - page

62.

5X00 Systems Troubleshooting

Issue

possible cause

possible solution

time and date lost at

power up

bad lithium battery

replace lithium battery

no events in event log

event logging not enabled

See Event Logging - page

112..

no data in data log

data log interval time has not

passed

See Data Logging - page

110.

cannot view specific data

log records at 5X00

Data log record has older

date and time-stamp than

the 25 most current data log

records available at 5X00

sensor log.

See Data Logging - page

110.

Firmware Updater Troubleshooting

Issue

possible cause

possible solution

Device “timed out” error incorrect configuration in

5X00 communication menu

See Device Timed Out er-

ror - page 122.

incorrect wiring of RS232

connection

See Device Timed Out er-

ror - page 122.

5X00 does not go into

downloader mode when

“program device” is

selected at Firmware

Updater

Manually put 5X00 in

downloader and retry firm-

ware updater upload.

Menus → System → Down-

loader
See Force downloader -

page 123..

Troubleshooting

Network Troubleshooting continued