5. Configure PC com port number (1-8) in Flash Programmer.

Notes:

- Use only a Tripp-Lite Keyspan USB-Serial Adapter, model USA-19HS YSI

p/n

773

.

- When using USB-Serial Adapter, verify that the port number assigned dur-

ing driver installation is the same port configured when using AquaManager

Flash Programmer.

6. AquaManager Program Folder includes the latest version of instru-

ment firmware. The lastest firmware is stored as an .mdb file in the PC

Program Files>AquaManager>Firmware. At AquaManager, Firmware

Updater>File>Load Firmware Database. Locate and load this file into

the Firmware Updater.

7. Left click

button.

window opens.

8. Select “Yes” and the following screens will appear.

Configuring the 5X00

On

instrument

On

PC

YSI 5200A/5400/5500 Bootstrap Loader screen

Flash Programmer uploading firmware into

5X00

Note:

- If the YSI 5200A/5400 Bootstrap Loader screen does not appear at the instru-

ment after above steps manually place 5X00 in downloader at Menus → System

→ Downloader. Close and reopen AquaManager. Reopen Flash Programmer

and repeat steps starting at 6.

- Power cycle to exit downloader and return to run screen. Do not power

cycle during an active upload.

- If after manually placing the 5X00 into downloader and the Flash Program-

mer utility still does not upload firmware see Force Downloader - page 123.

9. The 5X00 will reboot (power cycle) when firmware upload is complete.

10. Select “ok” when this window appears

11. Close Flash Programmer.

12. Verify version information to confirm downloaded code:

Menus → System → Version

13. If applicable, disable Force Serial Port configuration.