Input system specifications – YSI 5500D-1 User Manual

Page 8

Advertising
background image

YSI 5X00

15

YSI 5X00

14

Input System Specifications

Aux Systems

There are 6 Aux input systems that allow digital or analog (0-

1.0VDC, 0-5.0VDC, or 4-20mA) configuration. In addition, four

of the Aux systems can be configured as temperature systems see

Temperature Systems Specs - page 15.

Aux System Type Digital

Analog

Range

dry contact open/close

0 -1.0 VDC, 0-5.0 VDC, or 4-20mA

Accuracy

n/a

0-1.0VDC: ± 3% ± 10 mV

0-5.0VDC: ± 3% ±15 mV

4-20 mA: ± 3% ±100μA

Resolution

n/a

user selectable (0), (0 .0), or (0.00)

DO Systems

There are up to 4 independent DO systems. Each DO system has

2 control ranges, 2 control ranges, and high and low alarming.

Sensor Type

5400

Polarographic (mA)

Galvanic (mV)

Membrane Type Teflon (TE), 2 mil (gray)

Polyethylene (PE), 2 mil (blue)

Teflon (TE), 2 mil (gray)

Range

0 to 500% air saturation

0 to 60 mg/L

0 to 500% air saturation

0 to 60 mg/L

Accuracy

±2% or ± 0.2 mg/L, whichever is

greater for 0-200%

±6% of reading for 200-500%

±2% or ± 0.2 mg/L, whichever

is greater for 0-200%

±6% of reading for 200-500%

Resolution

0.1% (0.0-99.9%); 1% (> 99.9)

0.01 mg/L (0-10 mg/L); 0.1 mg/L

(10-60 mg/L)

0.1% (0.0-99.9%); 1% (> 99.9)

0.01 mg/L (0-10 mg/L); 0.1

mg/L (10-60 mg/L)

Response Time

95% of reading in 17 seconds for

2.0 mil PE membrane;

95% of reading in 36 seconds for

2.0 mil Teflon membrane.

Allow approximately 10-15 min-

utes after 5X00 is powered on, or

after the DO sensor is serviced,

for the polarographic sensor to

stabilize.

95% of reading in 36 seconds

for 2.0 mil Teflon membrane.

Cable lengths

4, 10, 20 m (special order up to

100m)

4, 10, 20 m (special order up to

100m)

Introduction

Flow require-

ments

minimum flow rate of 3 inches

per second or 7.62 cm/second for

2.0 mil PE membrane; 6 inches

per second or 15.24 cm/second

for 2.0 mil Teflon membrane

6 inches per second or 15.24

cm/second for 2.0 mil Teflon

membrane

Other

manual salinity compensation for

DO mg/L readings

manual salinity compensation

for DO mg/L readings

Sensor Type

5500D-01, 5500D-02,

5500D-04

Optical Dissolved Oxygen (dynamic luminescence quenching

lifetime detection)

Dimensions

19 cm length (7.5 in); 2.4 cm diameter (0.95 in)

Membrane Type

Sensor Cap includes calibration coefficients

Range

0-500% air saturation

0 to 50 mg/L

Accuracy

±0.2°C

Salinity Input Range

0-50 ppt

Storage Temperature

-20 to 70°C

Operating Tempera-

ture

-5 to 70°C

Response Time

90% in 25 seconds; 95% in 45 seconds

Cable lengths

1- 4- 10- 20- 30- 40- 50- 60- and 100-m;

(3.3, 13.1, 32.8, 65.6, 98, 131, 164, 197 and 328 ft)

Flow requirements

None

Temperature

Systems

There are 4 independent temperature systems. Auxiliary inputs

3, 4, 5, & 6 can be configured as temperature systems. Each tem-

perature system has 1 control range, 1 control  range, and high

and low alarming. Temperature systems can be configured as the

temperature source for DO measurements on a 5400 or as stand

alone systems on either 5400 or 5500 instruments..

Sensor type

10,000 ohm @ 25°C thermistor

Range

0-45°C; 32-113°F

Accuracy

± 0.2°C

Resolution

0.1°C

Introduction

Advertising