13. Enter DO calibration menu

Menus → Calibration→ DO→Sat

14. Allow approximately 10 minutes for the sensor to settle in the calibration

envrironment and for the temperature to equilibrate.

15. After 10 minutes, verify that the DO reading shows no significant change

for approximately 30 seconds. Then press

to complete calibration

OR press

to abort the calibration without saving.

16. Press

to save calibration and return to Run Screen.

17. Remove the probe from its calibration chamber.

18. Rinse sensor in tap or purified water and place the probe assembly back

in its monitoring location.

Directions to Calibrate DO Sensor mg/L

A large bucket can be used when calibrating the DO mg/L reading. Be sure to

provide sample flow across the membrane or sensor cap by physically stirring the

probe or using a stirrer such as a stir bar.

The DO sensor requires a sample flow rate of at least 3 inches (~7.62 cm) per second

if using the 2.0 mil PE membrane (blue) and at least 6 inches (~15.24 cm) per second

if using the 2.0 mil Teflon membrane (gray).

1. If you have not already done so, locate and install the sensor if using a

YSI 5422, see probe installation instructions included with the sensor.

2. Next, install a new membrane, or sensor cap, if necessary. See instruc-

tions included with membrane kit.

3. Verify that the sensor type and membrane type are configured correctly

in the DO Sensor Setup menu at the 5X00.

4. Read Getting Ready to Calibrate - page 83.

5. If using the 5422 polarographic sensor, wait at least 10-15 minutes after

the 5400 is powered on, or after the DO sensor is serviced, before cali-

brating. This allows enough time for the DO sensor to stabilize.

5400

5400

5400

Configuring the 5X00

6. Verify that the configured salinity (ppt) value in the DO Sensor Setup

menu is correct. The entered salinity value should equal the salinity of

the water being monitored (not the calibration solution). If necessary,

enter the correct the salinity value (ppt).

Menus → Sensor Setup→ DO→ppt

7. Place the DO probe in a water sample with a known DO concentration

(the DO mg/L value can be determined from another recently calibrated

DO instrument or from a Winkler titration).

8. If the temperature source for the DO sensor being calibrated is not com-

ing from a temperature sensor located on the DO sensor’s probe assembly,

make sure that the temperature value being reported by the source is

equivalent to the temperature at the DO sensor’s calibration location.

Menus → Sensor Setup→ DO→Temp Source (User Defined)

9. Enter DO calibration menu

Menus → Calibration→ DO→mg/l

10. Allow the probe to stabilize for approximately one minute.

11. After one minute, verify that the DO value shows no significant change

for approximately 30 seconds. Then, press

to complete the cali-

bration OR press

to escape calibration and return to calibration

menu without saving.

12.

Using the arrow keys to select numbers on the numeric keypad,

enter the known DO value in mg/L. Press

to confirm each number

selection.

13. Press

to confirm the entered value OR press

to abort

the calibration without saving any changes.

14. Press

to save the calibration and return to the Run Screen.

15. Rinse sensor in tap or purified water and place the probe assembly back

in its monitoring location.

16. Rinse sensor in tap or purified water and place probe assembly in system

stream.