Control menu, Msg’s menu – YSI 5500D-1 User Manual

Page 86

Advertising
background image

YSI 5X00

171

YSI 5X00

170

- The Run Screen must be displayed during normal operation. If the 5X00

is not at the run screen during AquaManager uploading, pressing

at

the 5X00 may only save certain configuration changes.

See Chapter 5 AquaManager - page 175 and AquaManager Online Help for ad-

ditional information.

Control Menu

Press

at the Run Screen to display a list of energized sensor and aux 5X00

system relays. The Control menu identifies the active sensor or aux system and

control and/or alarm condition. See Control mode and alarms - page 138 for

additional information.

Notes:

- Press

to display time stamped detail of control and alarm relay(s).

- Alarm message appears at the Control menu for aux and sensor systems when

alarms are enabled regardless if an alarm relay is assigned.

- Active timer (10 event timer and feed timer) systems are not displayed at

control menu.

- General alarm systems are not displayed at the Control menu regardless if

general alarm relay is enabled.

- Use

to scroll list when scroll bar appears at window.

When no 5X00 control or alarm system(s) are active, the below screen is displayed.

Msg’s Menu

Calibration, Sensor, and Event logs contain 5X00 system information. The 5X00

Sensor log stores 2000 records. The Calibration log stores 500 records. The Event log

stores 1250 records. Once the log memory is full, the oldest records are overwrit-

ten with more recent records. The number shown in ( ) at the Msg’s menu reflects

the total number of records in the log. The 25 most recent records for each log are
displayed at the

menu. Use

to scroll and press

to view time

stamped detail record.

Use AquaManager to capture all records from 5X00 logs and store them to a PC

database. Polling times should be determined by logging intervals. Example - if the

sensor log is recording values every minute and the sensor log holds 2000 records,

polling should occur at least every 33.3 hours. This allows for all records to be

saved to the PC before being overwritten on the 5X00. Set the data log interval so

that analyzing data based on the configured interval will be helpful. Poll at intervals

that don’t result in gaps in data.

A Reset Clean Probe message will appear at the Msg’s Menu if the clean probe timer

expires. Selecting “Yes” at “Are you sure?” resets the clean probe timer to the user

configured number of days.