Buzzer, Communication, Email alarming – YSI 5500D-1 User Manual

Page 46

Advertising
background image

YSI 5X00

91

YSI 5X00

90

1. Enter BackLight Timer menu

Menus → System →BackLight Timer

2.

Use

and

to scroll and highlight submenus.

3.

Press

to select the submenu you wish to change.

4. Enable and configure submenus.
5. After making desired changes, press

to save configuration and

return to Run Screen.

Buzzer

Menus → System → Buzzer

The 5X00 has an audible buzzer that sounds when a general alarm condition oc-

curs. The buzzer sounds after the alarm hold off period. The General Alarm must

be enabled for the buzzer to sound. Use the Buzzer Menu to enable or disable the

audible alarm.

When enabled, the buzzer will sound until one of the four bottom softkeys on the

5X00 front panel is pressed. From the Run Screen, press one of the following softkeys

to reset the General Alarm. Resetting the General

Alarm silences the buzzer, de-energizes any alarm relay(s), and stops any pending

5X00 email alarm messages. See General Alarm - page 124 and Sensor Setup - page

137 for additional information.

1. Enter Buzzer Menu

Menus → System → Buzzer

2. Press

or

to turn the buzzer either off or on.

3. Press

to save configuration and return to Run Screen.

Configuring the 5X00

Communication

Menus → System → Communications

Select the communication method for AquaManager and/or 5X00 email alarm-

ing. Only 1 communication method can be used at a time. Hardware changes are

required when the communication method is changed. See Wire Communication

Equipment - page 67. Configure remote ethernet communication, network set-

tings, and RS232 settings in the Communications menu.

Note:

- AquaManager supports both communication methods.

Email Alarming

Email alarming can be configured 1 of 2 ways.

1. A 5X00 email is sent from the 5X00 serial to ethernet device - See Ethernet

starting on page 93 for additional information.

2. An AquaManager email is sent from the PC running the AquaManager

application. AquaManager Autopoller must be running. See Chapter 5

AquaManager - page 175 and AquaManager online help for additional

AquaManager email alarming information.

*Ethernet connection requires access to an SMTP server accessible by the local

network.

EMAIL ALARMING DISCLAIMER - There are many factors that can affect deliv-

ery of an email in a timely fashion. Certain SMTP email servers are more reliable

than others. No email alarm message has guaranteed delivery. Use multiple email

addresses and email forwarding as safe guards. Mechanisms, such as spam filters

used by service providers, evolve constantly and can result in non-delivery of email

Configuring the 5X00