YSI 5500D-1 User Manual

Page 55

Advertising
background image

YSI 5X00

109

YSI 5X00

108

figure 4.7

Notes:

- Values displayed at the run screen are based on display format resolution

configuration. Example - if the display format is 0.0 and the min value input

is 1.88, then 1.9 will be displayed at the run screen.

- 5X00 aux analog (non temp) sensor log records 0.00 resolution regardless of

the selected resolution in the aux analog display format configuration.

Directions to configure Aux analog (non temp) Configure Aux menu:

1. Wire aux analog devices - page 59.

2. Enter Aux menu

Menus → System → Configure Aux →Aux 1-6

3.

Use

and

to scroll and highlight submenus.

4.

Press

to select.

5. Press

to save configuration.

6. Go to Sensor Setup Menus → Sensor Setup Aux 1-6 - page 149. to en-

able and configure set points, control, alarm, relays, and hold off time.

7. Press

to save configuration.

Note:

- The alarm low value must be ≥ the min value. The alarm high value must be

the max value.

Analog Temperature (Aux 3, 4, 5 and 6)

Menus → System → Configure Aux →Aux 3-6→Input Type→Temperature

Aux 3, 4, 5 and 6 can be wired and configured for a temperature sensor. Temperature

systems are configured differently for 5400 and 5500D instruments. The differences

are explained below.

5400 ~ Aux configured temperature systems can be configured as stand alone

5400

Configuring the 5X00

independent temperature systems. In addition, Aux temperature systems can be

configured as the temperature source for any of the four DO systems. Configure the

DO temperature source in the DO sensor set up menu - see page 154.

5500D-01, 5500D-02 and 5500D-04 ~ Aux configured temperature systems can

only be configured as stand alone independent temperature systems and cannot be

assigned as a temperature compensation source for any DO system. The DO system

temperature compensation value is always taken from the temperature sensor in the

ODO probe. Temperature systems can be configured using the ODO temperature

value or the Aux 3, 4, 5 and/or 6 temperature value. Configure the temperature

system source at the Menus → System → Configure Aux →Temp Source menu.

Configure “Aux Port” if the temperature system is to be driven by the input value

of a temperature sensor wired to aux 3, 4, 5 or 6. Configure “ODO Probe” if the

temperature system is to be driven by the ODO probe temperature sensor value.

Directions to configure Aux temperature Configure Aux menu:

1. Wire temperature sensors - page 54.

2. Enter Aux menu

Menus → System → Configure Aux →Aux 3-6

3. Configure temperature input type and name.
4.

Use

to scroll and highlight submenus.

5.

Press

to select.

6. Press

to save settings and return to Run Screen.

7. Go to Sensor Setup Menus → Sensor Setup Temperature for Aux 3-6

and ODO - page 150 to configure temperature system.

8. Press

to save settings and return to Run Screen.

Note:

- The Sensor Setup menu - page 137 is automatically enabled when an aux

channel is configured for temperature.

5500

5500

Configuring the 5X00

Advertising
This manual is related to the following products: