YSI 5500D-1 User Manual

Page 27

Advertising
background image

YSI 5X00

53

YSI 5X00

52

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5400 is po

w

er

ed

. S

er

ious damage can oc

cur

.

ground DO

sensors

Figure 3.19

Proceed to Step 5 following 5500 DO sensor wiring directions page 52 to complete

sensor wiring.

4. On 5500D instruments there are two terminal strips at location H. Wire

system DO-1 and system DO-2 sensors to the upper terminal strip. Wire

system DO-3 and system DO-4 sensors to the lower terminal strip - see

figures 3.19a and 3.19b. The upper and lower DO sensor terminal strips are

inverted. Be sure ODO sensor wires follow the wire coding below.

YSI 6060250 ODO/temperature digital sensor cable YSI
Wire Function

Color Coding

power/6.3 volts DC for ODO Probe

RED

GND/Sensor

GREEN

RS485 B/bidirectional data

WHITE

RS485 A/bidirectional data

BLACK

Ground/drain wire

BARE

• YSI 6060250 ODO DO sensor wires are red, green, white and black.

-Terminate red wire to 1R (“1”=DO1 “R”=red) and/or 2R, 3R, 4R.

-Terminate green wire to 1G (“1”=DO1 “G”=green) and/or 2G, 3G,

4G.

-Terminate white wire to 1W (“1”=DO1 “W”=white) and/or 2W,

3W, 4W.

5500

-Terminate black wire to 1B (“1”=DO1 “B”=black) and/or 2B,

3B, 4B.

-Terminate bare ground/drain wire to E or F on metal plate.

Notes ground/drain wire termination:

- Install a ring (eyelet) or spade (fork) terminal onto the ground bare wire before

terminating ground wire to E or F.

- Multiple ODO ground/drain wires can be terminated to the same ground

location on the metal plate.

- Return screw at E or F over terminal connection of ground wire.

Figure 3.19a

Figure 3.19b

Installation and Wiring

Installation and Wiring