Calibration logs, Event logs, Sensor logs – YSI 5500D-1 User Manual

Page 87: Svr hold softkey button, 173 f

Advertising
background image

YSI 5X00

173

YSI 5X00

172

Note:

- User configured sensor and relay names do not appear in log menus.

Calibration Logs

Calibration logs store sensor calibration data. Calibration logs are date and time

stamped and contain sensor and calibration information for completed calibra-

tions.

Certain calibration data is not available at the 5X00. Use AquaManager to view and

save calibration data to a PC. In addition to sensor system, time, date, and calibra-

tion method, AquaManager calibration logs include user input 1, user input 2, and

the original and new slope/offset information when applicable.

Event Logs

Control, Alarm, and Timer relays assigned to 5X00 systems create events when they

energize and de-energize. In addition, AC power fail, ethernet failure to initialize

errors and mode network failures create events. These events are recorded in the

Event log. Events identify the 5X00 system, are time stamped, and record relay on

and off times (if applicable). The General Alarm does not create an event or control

message unless a General Alarm relay is enabled.

Note:

- Enable event logging in the System menu. See Event Logging - page 112 for

additional information.

Sensor Logs

Enabled systems record sensor data based on data logging menu configuration.

Sensor logs identify sensor system, are date and time stamped, and record sensor

value at user configured intervals.

Notes:

- Data logging must be enabled in the System Menu. See Data Logging menu-

page 110 for additional information.

Svr Hold softkey Button

When servicing sensors or doing tank maintenance, it may be necessary to shut

off all systems (including relays) because sensor values will not be accurate. Press

at Run Screen to enter Service Hold menu. When in Service hold, all relays

are de-energized until the Service Hold time expires. The Service Hold time is user

selectable. Enable and configure service hold time at Menus → System→Service

Hold - page 134.. After service hold time expires or is cancelled, sensor and aux

Configuring the 5X00