Step 4 wire power, Wire ac power, Directions to wire ac power – YSI 5500D-1 User Manual

Page 24

Advertising
background image

YSI 5X00

47

YSI 5X00

46

7. Terminate other end of the earth ground wire to a cold water pipe, steel rod,

copper pipe, or earth ground.

8. Complete other wiring to configure 5X00 system.

9. Close front panel - page 43.

Step 4 Wire Power

The 5X00 is available as an AC or DC instrument. The DC version can only be

powered by DC power. The AC version can be powered by either AC or DC power.

If an AC version is properly wired and configured for both AC and DC power, and

AC power fails, the power switches to 12VDC. If DC voltage drops below a user

defined safe operating range, an alarm occurs.

CAUTION: At 120VAC, the 5X00 draws 25 watts (~0.2 amps). Use wiring

suitable for that load following all safety information and local electrical codes.

Back up power is recommended when operating with AC power. Backup power

can be supplied by 12VDC or UPS battery source.

Wire AC Power

The AC power option includes an integrated universal (worldwide) AC switching

power supply allowing AC power from any AC receptacle 100-240VAC. AC power

wires are terminated at location O on I/O Board - Figure 3.17. (Common lettering

references are used for both 5400 and 5500 models).

Directions to Wire AC Power

WARNING: Disconnect external power to the unit before wiring.

AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer

un câblage quelconque.

1. Perform steps 1-3 of 5X00 installation.

2. Open front panel - page 42

3. Feed AC power cable through drilled hole in rubber grommet to location

O. The terminal strip is removable.

CAUTION:

Run high and low voltage cables through separate bulkhead and

conduit.

4. Terminate AC power wires to pins “G” (ground), “L1”, and/or “L2/N” ac-

cording to applicable local electrical codes. Use minimum 18 gauge 3 wire

power cord.

CAUTION:

It is recommended to power only the 5X00 instrument with the power

cord to prevent exceeding the current draw of the AC power supply.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5400 is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.17

5. Complete other wiring to 5X00.

6. Close front panel - page 43.

7. Test AC power.

Test AC Power

Turn on main power. The enabled system values display on the Run Screen after

initial YSI splash screen. If applicable, install and configure DC power backup page

48. Verify AC power at the Menus → System →Version menu - page 135.

See power fail menu - page 128 for AC power event log information.

Installation and Wiring

Installation and Wiring