Relay test, Service hold, Version – YSI 5500D-1 User Manual

Page 68

Advertising
background image

YSI 5X00

135

YSI 5X00

134

Relay Test

Menus → System → Service→ Relay Test

Use the Relay Test menu to verify peripheral devices are wired correctly. Entering

this menu turns off any active relays and disables any sensor, aux, and timer system

relays from energizing regardless of value or condition. Systems are not reset when

the Relay Test menu is exited. Reset systems after exiting the Relay Test menu,

cycling power to the 5X00.

1.

Press

to toggle relay on.

2.

Press

to toggle relay off.

3. Press

to save configuration and return to Run Screen.

Note:

- Timer and control icon(s) remain at the Run Screen until the system is reset.

Example - the timer icon remains at systemwide location until the end of

the timer cycle that was stopped when the relay test menu was entered even

though the relay is not energized. To reset systems after exiting the Relay Test

menu, power cycle the 5X00.

Service Hold

Menus → System → Service Hold

Pressing

at the Run Screen puts the 5X00 into Service Hold. The Service

Hold time is user selectable (1-255 sec, min, or hours). Service Hold turns off any

active relays and disables any sensor, aux, and timer system relays from energizing

regardless of value or condition. 5X00 alarm emails are not sent during service hold.

After the service hold time expires or is cancelled, the sensor systems are reset after

30 seconds. Timer systems are not reset. See Service Hold Menu - page 173 for

additional information.

Configuring the 5X00

1.

Use

and

to scroll and highlight submenus.

2.

Press

to select.

3. Enable and configure submenus.
4. Press

to save configuration and return to Run Screen.

5. Press

at Run Screen to enter Service Hold . See Svr Hold page

173 for additional information.

Note:

- Place the 5X00 in service hold as instructed in step 3 of calibration checks in

Maintenance, page 189.

Version

Menus → System →Version

The Version menu contains information about the 5X00 including: display and

control board firmware code version, firmware compile date and time, control board

temperature, AC and DC voltage and model number.

Notes:

- Firmware version information is required when contacting YSI support.

- The clock chip temperature range starts at 0oC. Values < 0oC are not valid

numbers.

- The normal operating range temperature will be a few degrees above ambient.

- AC powered units will run at a higher temperature than DC powered units.
- To view current AC and DC power values

to return to run screen

and then return to menu. Power information does not refresh when version

menu is open.

Configuring the 5X00