Relay (outputs) troubleshooting, Email alarm troubleshooting – YSI 5500D-1 User Manual

Page 101

Advertising
background image

YSI 5X00

201

YSI 5X00

200

Issue

possible cause

possible solution

DO-galvanic or polaro-

graphic inaccurate values

DO sensor wires are termi-

nated incorrectly

see Wire DO sensors -page

50..

probe assembly ground wire

or temp sensor wires (if not

in use) are causing interfer-

ence)

cut back unused wires to

cable sheathing to avoid

shorting.

“over/under” sensor

readings

bad sensor

bad membrane

clean sensor, replace

membrane, calibrate sensor,

test sensor outside system

stream in solution of

known value,

replace sensor

no sensor value on

screen

sensor not enabled

See Sensor Setup - page

137..

display setting configuration

set for multiple Run Screens

See Display Setting - page

115.

“check probe” message at

sensor calibration

improper placement of probe

assembly, incorrect sensor

wire connection(s), improper

placement of 5X00, bad sen-

sor, and/or bad membrane

See Location Considerations

Probe Assembly - page 33

See Wire sensors - page

50.

See Location Consider-

ations 5X00 - page 27.

clean sensor, calibrate,

test sensor outside system

stream in solution of

known value, replace mem-

brane and/or replace sensor

if necessary

Analog aux value is

overlapping to adjoining

sensor window

1X1 & 2X2 display format

setting cannot display values

over 7 digits in aux sensor

window

Adjust aux display format

to lower resolution i.e. from

0.00 format to 0 format.

See also Notes: Display

Screen Limitations - page

116..

Troubleshooting

Sensor Troubleshooting continued

Relay (Outputs) Troubleshooting

- Disable control, alarm, and timer relays when troubleshooting relay out-

puts.

- Verify 5X00 wiring and configuration.

Issue

possible cause

possible solution

relay does not energize

incorrect sensor configura-

tion

See Sensor Set up - page

137.

relay not enabled

incorrect timer configuration See Timers - page 159..
30 second delay at power up

and/or sensor configuration

change

30 seconds delay - see page

228..

peripheral device is wired

incorrectly

See Wire Relays - page 56.

The device is incorrectly

wired to NO or NC termi-

nal strip ports..

relays are not enabled for

control/alarm/timer system

Verify menu configuration

(relay enable) for control/

alarm/timer system.

See sensor system set up -

page 137.

See timer system enable

-page 159.

temp (high) does not de-

energize when set point

value is read

sensor is configured for

timed control

Verify setpoint and control

configuration values - page

151.

Email Alarm Troubleshooting

Issue

possible cause

possible solution

no 5X00 email alarm

message received

5X00 display is not at Run

Screen - all alarming is sus-

pended when in menus

exit to run screen to reset

alarm email alarm system

Troubleshooting