Main menu, Calibration menu, Getting ready to calibrate – YSI 5500D-1 User Manual

Page 42

Advertising
background image

YSI 5X00

83

YSI 5X00

82

Main Menu

Press

at the Run Screen to enter the Main Menu.

All system configuration is done through this menu. The four Main Menu submenus

are Calibration, System, Sensor Setup, and Timers.

Calibration Menu

DO sensors(s) must be user calibrated to ensure accurate measurements. Calibrate

DO sensors at the Calibration Menu. Entering the Calibration Menu de-energizes

control and alarm relays for the DO system being calibrated. Energized DO control

relays for systems not being calibrated remain energized including any/all tempera-

ture systems regardless if they are the configured temperature (temp) source for

the DO sensor being calibrated (5400 only). If necessary, disable any temperature

relays for the associated DO sensor in the Sensor Setup menu. There is a 30 second
hold off period for control and alarm relays after pressing

or

at

the calibration menu. The hold off period allows sensors time to stabilize before

processing readings for controlling and alarming purposes.

Notes:

- Active Aux and Timer relays do not de-energize at calibration menu.

- A “Check probe” error message during calibration means that the sensor is

Configuring the 5X00

unable to calibrate within the 5X00’s specifications. This may indicate improper

sensor wiring (page 50), 5X00 location (page 27), sensor location or (page

33), the need to service the sensor, or a failed sensor.

- Calibration of temperature sensors is not required and is not available.

- Energized alarm relays for all systems de-energize when a front panel key

is pressed to enter the calibration menu. Alarm systems are not reset until

is pressed to return the instrument to the run screen.

Getting Ready to Calibrate

Calibrate DO sensor(s) by performing a one point calibration. Calibration can be

done to 100% Saturation (Sat) or mg/L. Calibrating to %Sat automatically calibrates

to mg/L and vice versa so it is only necessary to calibrate one or the other, not both.

For both ease and accuracy, YSI recommends calibrating in %Sat. Salinity com-

pensation for the mg/L readings is obtained from the user entered ppt value. This

value can be entered and modified in the DO sensor setup menu.

Figure 4.5

Figure 4.6

Configuring the 5X00