Email Retry Delay

Menus → System → Communications → Email Retry Delay

Configure the Email Retry Delay for the time interval that emails should be retrans-

mitted after the initial alarm email is sent. The initial alarm email is sent immediately

after the general alarm hold off time or after the aux hold off time. Emails will be

retransmitted until one of the front panel softkeys is pressed, regardless if the alarm

condition still exists.

Notes:

- On a network, emails are retransmitted according to the slave retry delay

configuration of the slave in alarm and not the master’s email retry delay

configuration. Please note, a slave Email Retry Delay menu will display (Not

available) because of its slave configuration; however, this feature is available

and has a default value of 15 minutes that can be configured by the user.

- The Retry Delay should be configured at intervals ≥ 15 minutes.

- AquaManager alarm emails are retransmitted based on 5X00 slave(s) and

master email retry delay configuration.

Baud Rate RS232

Menus → System → Communications → Baud Rate

Configure the Baud rate (for RS232 connections only) in the 5X00 to match the

baud rate configured in AquaManager. Use the default rate of 115200 for fastest

communication speed.

1. Enter Network menu

Menus → System → Communications → Network →Baud Rate

2.

Press

or

to scroll and highlight a selection.

3.

Press

to confirm selection.

4. Press

to save configuration and return to Run Screen.

Force Serial Port

Menus → System → Communications → Force Serial Port

Configuring the 5X00

Enable the Force Serial Port when the 5X00 is communicating via RS232 (direct)

to a PC and when the 5X00 has a serial to ethernet device.

1. Wire RS232 connection - page 68.

2. Enter Communications menu

Menus → System → Communications → Force Serial Port

3. Use the

softkeys to configure off or on.

4. Press

to save configuration and return to Run Screen.

Notes:

- RS232 connection is required to upload firmware using the Firmware Up-

dater.

- The RS232 icon

blinks at the Run Screen when in Force Serial Port mode.

Non-Networked (Stand Alone) 5X00s

Non-networked (stand alone) 5X00s must have the termination relay enabled. In

addition, non-networked (stand alone) 5X00s must be configured as a master in

the System Communication menu with no slave nodes enabled.

Configure AUX

Menus → System → Configure Aux 1-6

The 5X00 has six auxiliary (Aux) inputs. Aux inputs 1, 2, 3, 4, 5, and 6 can be con-

figured for digital or analog ( 0-1V,DC 0-5VDC, or 4-20 mA) inputs. In addition,

Aux 3, 4, 5, and 6 can be configured with a temperature sensor. Configure suffix,

minimum (Min), maximum (Max), Input Scale Min, Input Scale Max and display

format menus for non temperature analog inputs. To configure set points, control,

alarms and relays see specific sensor set up menus. See Aux digital sensor set up -

page 147. See Aux analog (non temperature) sensor set up - page 149. See Aux

and ODO temperature sensor set up - page 150.