Advertising
background image

YSI 5X00

61

YSI 5X00

60

See Digital Input example - figure 3.24 and Analog Input example - figure 3.25.

CAUTION: The auxiliary inputs are not isolated. Devices connected must be

electrically isolated from ground and the water.

D

o not c

onnec

t or disc

onnec

t IDC r

ibbon cable

when 5400 is po

w

er

ed

. S

er

ious damage can oc

cur

.

Figure 3.23

5. Complete other wiring to 5X00.

6. Close front panel - page 43.

7. Test aux systems.

Test Auxiliary Input Devices

Apply power to 5X00. See Aux Setup - digital page 148 and analog (non tempera-

ture) page 150 for configuration information. Verify display, control, and/or alarm

functionality based on user defined configuration.

Wire Aux Digital Input

1. Install aux digital device according to manufacturer’s instructions.

2. Connect switch wires to Aux Input 1 - 6. Connect ground wire to (G) at

location M

for aux 1 and 2 or to (-) at L

for aux 3, 4, 5, or 6. Connect the

other wire to (+) L

and to 1 or 2 at M

.

Instrument with
analog voltage
output

Auxiliary Input Connector

+ -

CAUTION: Do not input external voltage.

Figure 3.24

Wire Aux Analog Input

1. Install analog instrument according to manufacturer’s instructions.

2. Connect analog instrument wires to Aux Input 1 - 6. Connect ground wire to

the (G) at M for aux 1 and 2 or to (-) at L for aux 3, 4, 5, or 6. Connect the

other wire to 1 or 2 at M for aux 1 and 2 or to (+) at L for aux 3, 4, 5, or 6.

Instrument with
analog voltage
output

Auxiliary Input Connector

+ -

CAUTION: Observe correct polarity on analog input.

Figure 3.25