Power fail menu, Serial number, Service – YSI 5500D-1 User Manual

Page 65

Advertising
background image

YSI 5X00

129

YSI 5X00

128

-

displays at Run Screen when password is enabled.

Power Fail Menu

Menus → System → Power Fail

In the event of AC power failure, an AC version instrument’s power switches to

DC if properly wired. The switch from AC to DC power is automatic and does

not require 5X00 configuration. Enable Power Fail AC Monitor to generate event

when AC power fails and is restored. Because AC powered 5X00s may not reboot

properly after an AC power failure or AC power surge, it is highly recommended

to use a battery back system when 5X00 is powered primarily by AC power. See

page 196 for DC battery maintenance information.

Disabling the Power Fail System will disable AC power events and disable the General

Alarm in the event of low battery for DC power. Disable Power Fail AC Monitor

when powering 5X00 by a DC power source.

Enable Low Battery Monitor to trigger General Alarm in the event DC voltage is ≤

to low battery trigger. The Low Battery Alarm Hold Off time is five (5) seconds and

cannot be user configured. Once the low battery general alarm has been acknowl-

edged by pressing one of the 4 softkeys, it is not reset. See General Alarm - page 124.

1. Verify AC and DC power supplies are properly wired and powered See

wire power - page 46.

Configuring the 5X00

2. Enter Power Fail Alarm menu and submenus

Menus → System → Power Fail

3. Use

to scroll and highlight submenus.

4.

Press

to select.

5. Enable and configure submenus
6. Press

to save configuration and return to Run Screen.

Notes:

- Press

to view AC power fail events.

-

icon at Run Screen indicates that the 5X00 DC power is ≥ Low Battery

Trigger value with Low Battery Monitor enabled.

-

icon at Run Screen indicates DC voltage ≤ the Low Battery Trigger when

Low Battery Monitor is enabled.

- Low battery alarm is not reset when front panel softkey is pressed.

- AC power fail does not generate a general alarm.

Serial Number

Menus → System → Serial Number

The serial number is displayed at the serial number menu. The serial number is

required when contacting YSI customer service. The serial number is also neces-

sary to configure AquaManager. 5400 instruments begin with “54-”followed by

seven digits. 5500D-01, 5500D-02 and 5500D-04 instruments begin with “55-”

followed by seven digits. Serial numbers are generated at the factory and cannot be

changed by the user.

Service

Menus → System → Service

The Service menu includes submenus to configure and Reset Clean Probe timer,

Configuring the 5X00