YSI 5X00

174

systems are reset causing relays to energize/de-energize based on the system value/

condition. Timer systems are not reset. To reset timer systems, turn power off and
then on to the 5X00. Press

to exit Service Hold.

The Run Screen displays a clock count down of the Service Hold remaining time.

Notes:

- If a timer is active (relay energized) at service hold, the relay will not reenergize

after service hold expires regardless if timer time has not expired. There is

only one “on” time for all timer systems.

- Entering Downloader mode places 5X00 into Service Hold.

- “Inst. Service Hold” is displayed on any AquaManager map icon associated

with an instrument in Service Hold. AquaManager alarm emails are not sent

for any alarm condition for a 5X00 that is in Service Hold.

- 5X00 alarm emails are not sent during service hold.