Yokogawa DR240 User Manual

Page 150

Advertising
background image

10-8

IM DR231-01E

10.3 Select Alarm Interval/Hysteresis/Hold/A/D Converter Integration Time/Filter

Selection of Interval for High Limit of Rate-Of-Change Alarm (ALARM RH TIME)

Set the number of measurement repeats in “Interval = measurement interval x measurement
repeats.”
• Select 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14, or 15.
• The default setting is 8.

Selection of Interval for Low Limit of Rate-Of-Change Alarm (ALARM RL TIME)

• The same selection method as the above.

Selection of Alarm Hysteresis (ALARM HYS)

Applied to high/low limit alarm.
• Select 0.0, 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9 or 1.0%.
• The default setting is 0.5%.

Selection of Alarm Display Hold (ALARM HOLD)

Select whether the alarm display is held or not when the alarm is reset.
• OFF: hold
• ON: not hold
• The default setting is OFF.

Selection of A/D Converter Integration Time (A/D INTG)

Select the A/D converter integration time. Select it from the following:
• AUTO: the frequency of 50/60 Hz is automatically switched corresponding to the power

frequency of this instrument. However, AUTO does not function when using DC power
supply. Selecting AUTO will set the A/D integration time to 50 Hz (20 ms).

• 50 Hz: the integration time is set to 20 ms (50 Hz).
• 60 Hz: the integration time is set to 16.7 ms (60 Hz).
• 100 ms: the integration time is set to 100 ms (10 Hz)
• The default setting is AUTO. If the model is a DR130/DR231 with /P6 option or a DR241 with

-2 power supply voltage specification, it is set to a default value of 50 Hz (20 ms). If you are
using the instrument on a 60-Hz power supply, change the A/D integration time to 60 Hz (16.7
ms).

Selection of Filter (FILTER)

Select whether the input filter is operated or not. Select it from the following:
• OFF: Not operated
• ON: Operated
• The default setting is OFF.

Note

• In order to make the set or selected contents effective, the contents must be registered. For details, see

page 10-24.

• To return from the SET UP menu to the operation display mode, select END in the main menu. For

details, see page 10-24.