13-10

IM DR231-01E

13.5 Calibration

Checking calibrated values in the DISPLAY mode

Use this mode only for checking calibrated values. In this mode, you can adjust a calibrated value,
but, if this value is invalid, the instrument may be inoperative. For details, contact your nearest
Sales representative.

11. With

, select the DISPLAY mode.

AD JUST MODE=DISPLAY

>MODULE No=0

DISPLAY CAL/EXEC END

12. Press the ENTER key to display the calibrated value display items.

The following shows where the universal input modules are calibrated.
The calibrating procedures for DC V/TC/DI input modules are those except RTD in universal
input modules.
13. The calibrated values are displayed as follows:

20mV ZERO=–

1

>MODULE No=0

>

14. Press the ENTER key.
15. Display the calibrated-value display items in the following order:

Repeat steps 13 and 14 above.
20 mV SPAN

→ 60 mV ZERO → 60 mV SPAN → 200 mV ZERO → 200 mV SPAN → 2 V

ZERO

→ 2 V SPAN → 6 V ZERO → 6 V SPAN → 20 V ZERO → 20 V SPAN → 50 V

ZERO

→ 50 V SPAN → Pt: 1 mA ZERO → Pt: 1 mA SPAN → Pt: 2 mA ZERO → Pt: 2 mA

SPAN

→ Pt: 1 mA-H SPAN → Pt: 2 mA-H SPAN → Cu: 2 mA ZERO → Cu: 2 mA SPAN

16. After checking the [Cu: 2 mA SPAN] calibrated value in the above calibrated-value display

items, press the ENTER key.

* * * DISPLAY END * * *

>MODULE No=0

>

17. Press the ENTER key.