13.4 Error Codes

100

No equation option

Not possible to make any computation
settings.

101

Invalid code in the equation

Set the equation correctly.

102

Incorrect number of ( ) in the equation

Set the correct number of ( ).

103

Syntax error

Set the equation correctly.

104

Illegal character(s) or code(s) is used before

Set the equation correctly.

or after an operator.

105

The right and left spans are the same.

Change the span.

106

Attempted to make range setting during

Stop making range setting.

computation.

110

Syntax error for computation constant

Set the constant correctly.

111

Out of constant setting range

Set the constant within the range.

120

Floppy disk not inserted or not formatted.

Insert a formatted floppy disk into the
drive.

121

Insufficient memory

Delete unnecessary files.

122

Incorrect file name or volume name

Set the correct file name or volume name.

123

Attempted to make settings

Stop making the settings.

which are not allowed to be made
during saving or reading of measured data.

124

Too many files in the media

Delete unnecessary files.

125

Attempted to read data which was saved

Select data which has been saved

by another instrument model.

by DR/DA.

127

Attempted to save data in a write-protected file.

Cancel write-protection.

128

No event/action has been set.

Set the desired action .

130

Incorrect channel found during saving

Set the correct channel.

or reading of measured data.

132

Attempted to format the disk while a file is open.

Close the file.

134

Set-up data saving error

135

Internal data error

136

Incorrect copy channel

Set the correct copy channel.

139

Other media related error

141

Disconnected modules were selected.

Specify correct modules.

143

Modules selected which cannot be calibrated.

Specify correct modules.

144

Calibration done improperly.

Check electrical connections.

Calibrate again.

If the same error occurs again, contact
your nearest Sales representative.

145

Hardware becomes faulty during calibration.

Turn off the power and then on.
Calibrate again.
If the same error occurs again, contact
your nearest Sales representative.

170

The IP address does not belong to any of the

Set the correct IP address.

classes, A, B, or C.

171

There is a hole in the mask or the host address

Set the correct subnet mask.

section is not released.

172

The net address section including the subnet

Set the correct net address.

does not match the subnet section of the IP
address.

173

The host address section of the IP address is

Set the correct IP address.

either all zeroes or all ones.

ROM ERROR

System ROM error

Contact your nearest sales
representative.

ROM ERROR*

System ROM error

Contact your nearest sales
representative.

RAM ERROR

Main memory error

Contact your nearest sales
representative.

RAM-DISK ERROR

RAM disk error

Contact your nearest sales
representative.

FLOPPY ERROR

Floppy disk error

Contact your nearest sales
representative.

MAIN NV READ ERROR

Internal nonvolatile memory read error

Contact your nearest sales
representative.

MAIN NV WRITE ERROR

Internal nonvolatile memory write error

Contact your nearest sales
representative.

CARRIAGE CANNOT MOVE

Printer eror

Contact your nearest sales
representative.

SYSTEM ERROR

System error

(

: number)

Check whether each unit or

(

: number)

power code is connected
properly after turning off the
power switch. If the same
error occurs again even though
you turn on the power switch,
contact your nearest sales
representative.

RESET ERROR

An error occurs between the main unit

Same as above.

and communication module.