13.5 Calibration

Entering the calibration mode

1. While pressing and holding the MODE key, turn on the power.

Selecting the module numbers

2. With

, select the module number.

MODULE No=0

0 1 2 3 4 5

3. Press the ENTER key.

Calibrating in CAL/EXEC mode

4. With

, select the CAL/EXEC mode.

AD JUST MODE=CAL/EXEC

>MODULE No=0

DISPLAY CAL/EXEC END

5. Press the ENTER key.
6. With

, select the calibration range.

AD JUST RANGE=20mV

>MODULE No=0

20mV 60mV 200mV 2V 6V 20V 50V Pt:1mA P

7. Enter the rated value in the selected range.
8. Press the ENTER key to execute calibration. During the calibration, the following display

appears.

A/D Adjusting

>MODULE No=0

20mV 60mV 200mV 2V 6V 20V 50V Pt:1mA P

9. After the calibration is completed, the following display appears.

* * * CALC END * * *

>MODULE No=0

>AD JUST RANGE=20mV

10. Press the ENTER key.