Yokogawa DR240 User Manual

Page 94

Advertising
background image

5-5

IM DR231-01E

Setting the Input Type/Recording Span/Linear Scaling

5

001-01:SCL:RTD/PT1

001-01:SCL:RTD/PT1

PT1 PT2 JPT1 JPT2 PT50 NI1 NI2 NI3

CU1 CU2 CU3 CU4 PT1S PT2S JPT1S

JPT2S J263B

SPAN=–_200.0/_ _600.0˚C

>Span limit(–200.0~600.0˚C)

SCL=_ _ _0.00/_100.00

***SET OK***

001-01:SCL:DI/LEVL

001-01:SCL:DI/LEVL

LEVL CONT

SPAN=

_ _ _ _ _ _

0/

_ _ _ _ _ _

1

>Span limit(0~1)

SCL=_ _ _0.00/_100.00

***SET OK***

ENTER

ENTER

ENTER

ENTER

ESC

ENTER

ESC

ENTER

From the previous page

The setting of the channel number is the same as explained in 5.1 on page 5-2. The following
explanation assumes that you already carried out the channel setting and that you selected “SCL”
as the input type.

Selecting the Input Type for Linear Scaling

The input type can be selected from the following. The default setting is VOLT.
• VOLT (DC voltage)

Refer to page 5-3.

• TC (thermocouple)

Refer to page 5-3.

• RTD (resistance temperature detector)

Refer to page 5-3.

• DI (contact)

Refer to page 5-3.

Setting the Recording Span

Refer to page 5-3.

Setting the Scaling Values (SCL)

The left scaling and right scaling values are set following the left and right span values of the span
menu. The value on the left side of the SCL menu shows the left scaling value, and the value on
the right side of the SCL menu shows the right scaling menu. The setting ranges from -30000 to
30000.
The decimal point can be set in any position of the scale as shown below. Set it when the left
scale is set.
" .

," "

.

," "

.

," "

. ," or "

."

• The default settings are 0.00 for the left span and 100.00 for the right span.
For details on setting the engineering units for linear scaling, refer to page 6-2.

5.2 Setting Linear Scaling and the Recording Span