Yokogawa DR240 User Manual

Page 241

Advertising
background image

14-14

IM DR231-01E

Installation height

Altitude up to 2,000 m

Installation category based on IEC 1010-1

II*1

Pollution degree based on IEC 1010-1

2*2

Warm-up time

At least 30 minutes after power switch-on.

*1Installation category is the specification of the inpuls withstanding voltage which is also called as

overvoltage category.

*2Polution degree is the level of foreign body adhesion such as the solid, liquid, and gas which decrease the

withstanding voltage, 2 means general indoor atmosphere.

Effect of Operating Conditions

Ambient temperature: Variation for a temperature change of 10

°C: within ±(0.1% of rdg + 1

digit);

±(0.2% of span + 1 digit) for Cu10Ω

Voltage variation: power supply: within

±1 digit over the range of 90 to 132, or 180 to 250 VAC

(AC power supply, frequency 50/60Hz), 10 to 32 VDC (DC power supply)
External magnetic field: variation with respect to AC (50/60Hz) and DC magnetic fields of 400
A/m: within

± (0.1% of rdg + 10 digits) Except for power monitor module: within ±15% of range

Signal source resistance: variation with respect to signal source resistance 1 k

Ω change

(1) voltage

2 V range or below: within

±10µV

6 V range or above: within 0.1% of rdg

(2) thermocouple

within

±10µV; however it must be within 100µV when burnout is specified.

(3) RTD

Variation with respect to change of 10

Ω per wire (when all wires have the same resistance

value)
Indication: within

±(0.1% of rdg + 1 digit)

Variation in indication with respect to a difference of 40m

Ω in the resistance between

conductors (max. difference between 3 wires): approx. 0.1

°C

Mounting position
Variation when the unit is mounted horizontally on a panel: within

±(0.1% of rdg + 1 digit)

Vibration
Variation when sinusoidal vibration of acceleration 0.2m/s

2

is applied for two hours in each of the

3 axial directions over a frequency range of 10 to 60Hz: within

±(0.1% of rdg + 1 digit)

Transportation and Storage Conditions

These refer to the environmental conditions existing during transportation and storage from the
time of shipment from the factory until commencement of use, and also during transportation and
storage in the case of a temporary period of non-use.
If the environmental conditions are specified within the specified range, the unit will not incur
permanent damage, and can be returned to a normal working condition (although re-adjustment
may be required in some cases).

Ambient temperature

–25 to 60

°C (-20 to 60°C with floppy disk drive)

Humidity

5 to 95% RH

Vibration

10 to 60Hz 4.9m/s

2

max

Shock

392 m/s

2

max (in packed condition)

14.1 Specifications of DR130/DR231/DR241 (Style S3)

Advertising