5 connecting the signal lines, Connecting the signal lines -16, Wwarning – Yokogawa DR240 User Manual

Page 57: Caution

Advertising
background image

3-16

IM DR231-01E

3.5 Connecting the Signal Lines

W

WARNING

• To prevent electric shock always make sure that the power supply is

turned OFF before connecting.

• When 30VAC or 60VDC and more is applied to the output terminal of

the alarm module or the output terminal of the DI/DO module, use
double-insulated wires(withstand voltage performance: more than
2300VAC) for those wires which apply 30VAC or 60VDC and more. All
other wires can be basic-insulated(withstand voltage performance:
more than 1350VAC). Furthermore, use “crimp-on” lugs (for 4mm
screws) with insulation sleeves for connecting to the screw terminal.
Make sure that the crimp-on tool must be one specified by the crimp-on
lugs manufacture, and that the crimp-on lugs and tool must be matched
to the wire size. To prevent electric shock, do not touch the terminal
after wiring and make sure to re-apply the cover.

CAUTION

• Do not apply an input voltage exceeding the following levels to each

terminal of each module. Otherwise, the internal circuits may be
damaged.
Allowable input voltage

Universal or DCV/TC/DI input module

2 VDC range or less, RTD, TC and DI(CONT) :

±

10 VDC

6 to 20 VDC range, DI(LEVEL) :

±

60 VDC

DI/DO module : –2 to 7 VDC

Max. common mode noise voltage

Universal or DCV/TC/DI input modules : 250 VACrms (50/60 Hz)

• Output contact rating for DI/DO or Alarm input modules is 250 VDC/0.1

A (resistive load), 250 VAC/2 A (resistive load), 30 VDC/2 A (resistive
load).

• The overvoltage categoly of each input module is CAT ll (IEC 1010).
• When connecting to a clamp terminal, use a signal conductor with the

following cross-sectional width:

Solid conductor

0.14 to 2.5 mm

2

Stranded conductor

0.14 to 1.5 mm

2

AWG

26 to 14

1 Check that the power switch of this instrument is turned off.
2 Remove the terminal cover. (the figure below shows DU100-11.)

Terminal cover

Screws for fastening the cover

3 Fasten the signal wires to the terminals as shown in the figure on the next page.
4 Re-apply the terminal cover and fasten the screws.

Advertising
This manual is related to the following products: