14-3

IM DR231-01E

Specifications

14

14.1 Specifications of DR130/DR231/DR241 (Style S3)

Measurement interval

DR130 : Selectable from 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 or 60s; max. 2s/20ch (measured with
20ch, filter:OFF, A/D integration time: 20ms(50Hz)/16.7ms(60Hz))
DR231/DR241 : Selectable from 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 or 60s; max. 2s/30ch (measured
with 30ch, filter:OFF, A/D integration time: 20ms(50Hz)/16.7ms(60Hz))

Noise rejection

Rejection by integration type A/D, lowpass filter, or moving averaging.

Input resistance

Min. 10M

Ω at 2V DC or lower and thermocouple range

Approx. 1M

Ω at 6V DC or higher.

Insulation resistance

Min. 20M

Ω at 500V DC between the input terminal and ground.

Input bias current

max. 10nA

Dielectric strength

Between input terminals

: 1000V AC (50/60Hz) for one minute

Between an input terminal and ground : 1500V AC (50/60Hz) for one minute

Input source resistance

DCV, TC : 2k

Ω or lower

RTD

: 10

Ω or lower per line (Pt100Ω)

5

Ω or lower per line (Pt50Ω)

1

Ω or lower per line (Cu10Ω)

Temperature coefficient

zero

: 0.01% of range/

°C

full span

: 0.01% of range/

°C (0.02% of span/°C for CU10Ω)

Thermocouple burnout

Detected in a TC range (On/Off) enabled, current of 4

µA, detectable pulse width of approx. 5ms

2 k

Ω or lower is considered “normal”

100 k

Ω or greater is considered “disconnected”

Detection interval for thermocouple burnout

Detected at each measurement interval.

Input Type

Power monitor

Terminal Configuration

Single-phase use (one channel each for voltage and current): clamp
Three-phase use (three channels each for voltage and current): clamp

Shortest Measurement Interval

2 s

Method of Input

Transformer-isolated input, with isolation between channels (separated channels)

Method of Measurement and Computing

Digital multiplication

Input Ranges

Voltage: 25 or 250 Vrms
Current: 0.5 or 5 A
In three-phase measurement or single-phase three-wire measurement, the current/voltage ranges
are identical between the respective phases or lines (the current/voltage ranges are set in common
between the respective channels).

Measured Frequency Range

45 to 65 Hz

Wiring Methods

DU400-12: Single-phase two-wire
DU400-22: Single-phase two-wire, single-phase three-wire, three-phase three-wire (dual-voltage,
dual-current), three-phase three-wire (triple-voltage, triple-current) and three-phase four-wire