Specifications

14

General Specifications

EMC Conformity Standard

• Emission

EN55011:Class A

• Immunity

EN50082-2

IEC1000-4-2

Electrostatic Discharge

Performance Criteria A*

8kV(Air), 4kV(Contact)

IEC1000-4-3

Radiated fields

Performance Criteria A*

80~1000MHz, 10V/m

IEC1000-4-4

Fast Transients

Performance Criteria B

Power line 2kV, The others 1kV

IEC1000-4-6

Conducted Disturbance

Performance Criteria A*

0.15~80MHz, 10V

* Effect on accuracy

±10% of range (except for 50V range)

±20% of range (50V range)
When the extender module is used:
±20% of range (except for 50V range)

±40% of range (50V range)

Installation

Desk-top type: DR130/DR231
Panel-mount type: DR241

Regardless of which installation method you use, be sure to install the units in an upright
position.

Materials

Steel plate, aluminium plate, plastic moldings

Color

Display: Slate Gray light (equivalent to Munsell 0.1 PB 4.6/0.2)
DR130/DR231 main unit: Lamp Black (equivalent to Munsell 0.8 Y 2.5/0.4)
DR241 main unit: Ice White (equivalent to Munsell 6.6 Y 7.9/0.5)

External dimensions

DR130 : approx. 338 (W)

× 221 (H) × 335 (D) mm

DR130 (with /P6 option) : approx. 338 (W)

×221 (Η) ×381 (D) mm

DR231: approx. 438 (W)

× 291 (H) × 336 (D) mm

DR231 (with /P6 option) : approx. 438 (W)

×291 (Η) ×381 (D) mm

DR241: approx. 444 (W)

× 288 (H) × 343 (D) mm

Weight

DR130: approx. 9.3kg

*1

, 10.8kg

*3

DR231: approx. 13kg

*2

, 14.5kg

*3

DR241: approx. 16kg

*2

*1 with 20 input channels and alarm output (/A4) installed
*2 with 30 input channels and alarm output (/A4) installed
*3 with /P6 option

Clock

with calendar function

Clock accuracy

±100ppm. However, this does not include the delay (less than 1 second) caused when the power is
switched ON and OFF once.

Fail Output

Contact output (when /R1 option is selected)

Set value backup

Lithium battery backup (approx. 10 years, at ambient temperature of 23

°C)

Insulation resistance

At least 20M

Ω between the power supply and ground, between each terminal and ground, and

between each input terminal (measured with 500 VDC)

Dielectric Strength

Between power supply terminal and ground of DR130/DR231/DR241 main unit: 1500 VAC (50/
60Hz) for one minute, Between input terminal and ground of DR130/DR231/DR241 main unit:
1500 VAC (50/60Hz) for one minute, Between output terminal and ground of DR130/DR231/
DR241 main unit: 2300 VAC (50/60Hz) for one minute, Between DC power supply terminal and
ground of DR130/DR231 main unit: 500 VAC (50/60Hz) for one minute (with /P6 option),
Between DC power supply terminal and ground of DR241 main unit: 1000 VAC (50/60Hz) for
one minute (in case of power supply code -2).

14.1 Specifications of DR130/DR231/DR241 (Style S3)