Yokogawa DR240 User Manual

Page 240

Advertising
background image

14-13

IM DR231-01E

Specifications

14

Power Supply

Rated supply voltage

AC power supply: 100 to 240 VAC (freely selected)
DC power supply: 12 to 28 VDC
AC power supply for /P6 option: 100 to 240 VAC (100/200 V auto switch)
Except, if the instrument is operated at 200 V it does not switch back to 100 V. In this case, turn
OFF the AC power switch on the back panel, and turn it back ON again.

Permissible supply voltage

AC power supply: 90 to 250 VAC
DC power supply: 10 to 32 VDC
AC power supply for /P6 option: 90 to 132/180 to 250 VAC

Rated supply frequency

50/60 Hz

Power consumption

AC power supply: Max. approx. 130 VA
DC power supply: Max approx. 80 VA
AC power supply for /P6 option: Max. approx. 130 VA

Fuse Ratings

• Main unit

AC power supply:
Maximum rated voltage/current: 250 V/2.5 A, Type: Time-lag, Standard: IEC/VDE
DC power supply:
Maximum rated voltage/current: 250V/6.3A, Type: Time-lag, Standard: IEC/VDE

• In the case at the rear side of the main unit where the input/output module is connected.(The

fuse may not be replaced by the customer.)
AC power supply:
Maximum rated voltage/current: 250 V/2 A, Type: Time-lag, Standard: IEC/VDE
DC power supply:
Maximum rated voltage/current: 250V/6.3A, Type: Time-lag, Standard: UL/CSA
AC power supply:
Maximum rated voltage/current: 250V/2.5A, Type: Time-lag, Standard: IEC/VDE

Priority on the use of the AC and DC power supplies on models with the /P6 option
(when both AC and DC powers are supplied.)

DC power supply terminal voltage

Power supply input with higher priority

Less than 20 V

AC power supply

20 to 28 V

Indefinite

28 to 32 V

DC power supply

Normal Operating Conditions

Supply voltage

AC power supply: 90 to 250 VAC
DC power supply (in case of DR130/231 with /P6 option or DR241 with -2 power supply cord):
10 to 32 VDC
AC power supply for /P6 option (for DR130/DR231 only): 90 to 132/180 to 250 VAC

Supply frequency

50Hz

±2%, 60Hz ±2%

Ambient temperature

0 to 50

°C (5 to 40°C when using floppy disk)

Ambient humidity

20 to 80%RH for 0 to 40

°C, 10 to 50%RH for 40 to 50°C (no condensation)

Vibration

10 to 60 Hz, 0.2m/s

2

Shock

Not allowed

Magnetic field

400 A/m max. (50/60Hz)

Position

Unit should be positioned left-right horizontally

Installation location

Room

14.1 Specifications of DR130/DR231/DR241 (Style S3)