Ad just mode=end, A/d adjust end=abort – Yokogawa DR240 User Manual

Page 226

Advertising
background image

13-11

IM DR231-01E

Trouble-shooting

and Maintenance

13

13.5 Calibration

Completing the selected module calibration with the END mode

18. With

, select the END mode.

AD JUST MODE=END

>MODULE No=0

DISPLAY CAL/EXEC END

19. PRESS the ENTER key.
20. With

, select STORE or ABORT.

A/D ADJUST END=ABORT

>MODULE No=0

ABORT STORE

21. Press the ENTER key. Return to operation 2.

Terminating the calibration mode

22. When you return to normal measurement after completing calibration, turn off the power and

then back on.

CAL/EXEC mode (for calibration of a specified range)
• Items for calibration range

• Items for the calibration range vary depending on the type of module to be calibrated.
• For use with universal input modules, select the items for the calibration range from among

the following:
20 mV/60 mV/200 mV/2 V/6 V/20 V/50 V/Pt: 1 mA/Pt: 2 mA/Pt: 1 mA-H/Pt: 2 mA-H/Cu:
2 mA

• When the DC V/TC/DI input module is selected, RTD in the universal input module is not

displayed.

• Displaying error messages

The following error messages are displayed.

• Attempted to remove the module during calibration. Or, hardware error: ERROR145
• Invalid calibrated data: ERROR144

Note

• When calibrating a high-precision RTD (Pt: [ ]-H), do so after calibrating the voltages to maintain

accuracy.

• When a calibrated data error appears, the error data are stored in memory.

• If ERROR145 appears, turn off the power and then back on to restart operation. If the same error occurs

again, the recorder may be defective.

• If ERROR144 appears, check the electrical connections and input data. If the same error occurs again,

the module to be calibrated may be defective.