Setting the Input Type/Recording Span/Linear Scaling

5

Setting the Recording Span (SPAN)

Set the left and right spans within the limits of an input range. In the SPAN menu item, the left-
hand value is the left span and the right-hand value the right span. Set the recording span within
the measuring range. The measurable limits vary depending on the measuring range you select,
as shown below.

25V-0.5A

25V-5A

250V-0.5A

250V-5A

Effective voltage

0.00 to 25.00V rms

0.00 to 25.00V rms

0.0 to 250.0V rms

0.0 to 250.0V rms

Vi (i=1,2,3,13,0)

Effective current

0.0000 to 0.5000A rms 0.000 to 5.000A rms

0.0000 to 0.5000A rms

0.000 to 5.000A rms

Ii (i=1,2,3,13,0)

Active power

–12.50 to 12.50W

–125.0 to 125.0W

–125.0 to 125.0W

–1250 to 1250W

P1,P2,P3

Active power

–25.00 to 25.00W

–250.0 to 250.0W

–250.0 to 250.0W

–2500 to 2500W

P13

Active power

–37.50 to 37.50W

–375.0 to 375.0W

–375.0 to 375.0W

–3750 to 3750W

P0

Apparent power

0.00 to 12.50VA

0.0 to 125.0VA

0.0 to 125.0VA

0 to 1250VA

VA1,VA2,VA3

Apparent power

0.00 to 25.00VA

0.0 to 250.0VA

0.0 to 250.0VA

0 to 2500VA

VA13

Apparent power

0.00 to 37.50VA

0.0 to 375.0VA

0.0 to 375.0VA

0 to 3750VA

VA0

Reactive power

0.00 to 12.50Var

0.0 to 125.0Var

0.0 to 125.0Var

0 to 1250Var

Var1,Var2,Var3

Reactive power

0.00 to 25.00Var

0.0 to 250.0Var

0.0 to 250.0Var

0 to 2500Var

Var13

Reactive power

0.00 to 37.50Var

0.0 to 375.0Var

0.0 to 375.0Var

0 to 3750Var

Var0

Power factor

–1.00 to 1.00

–1.00 to 1.00

–1.00 to 1.00

–1.00 to 1.00

PFi(i=1,2,3,13,0)

Phase

–80.0 to 80.0deg

–80.0 to 80.0deg

–80.0 to 80.0deg

–80.0 to 80.0deg

PHi(i=1,2,3,13,0)

Frequency

45.00 to 65.00Hz

45.00 to 65.00Hz

45.00 to 65.00Hz

45.00 to 65.00Hz

FREQ

Precautions in Measurement

• Input the voltage or current being measured at a level between 10% and 100% of the measuring

range. The DR recorder bases its calculations of all other parameters on the frequency of V1.
Extra care must therefore be taken when setting the input level of V1. If the input level fails to
fall within the given limits, there is no guarantee that the measurement of any other parameters
will be reliable.