2 specifications of optional functions, Specifications of optional functions -16 – Yokogawa DR240 User Manual

Page 243

Advertising
background image

14-16

IM DR231-01E

14.2 Specifications of Optional Functions

GP-IB Communication (when equipped with /C1)

Electrical and mechanical specifications

conform to IEEE Standard 488-1978

Code

ISO (ASCII) code

Address

0 to 15

Functions

• Talker functions

Output of measurement values (ASCII, binary); output of setting parameters (ASCII)

• Listener functions

Setting of measurement conditions, controls of starting and stopping measurement, specifying
causes of interrupts (excluding setting and control of power ON/OFF)

Causes of interrupts

Syntax error, chart end, completion of A/D conversion, operations of internal timer, end of data
saving/reading, incomplete measurement during computation

RS-232-C Communication (when equipped with /C2)

Electrical and mechanical specifications

conform to standard EIA RS-232-C

Connection method

point-to-point

Communication method

half-duplex

Synchronization mode

Start-stop synchronization (synchronized by a start and a stop bit)

Baud rate

150, 300, 600, 1200, 2400, 4800, 9600, 19200, 38400 bps

Start bit

Fixed at 1 bit

Data length

7 or 8 bits, selectable

Parity

Selectable from even, odd, or none

Stop bit

1 or 2, selectable

Transmission distance

max. 15 m

Connector

D-sub 25pin

Handshake

hardware : transmission and reception control by ‘DTR’’ ‘RTS’, ‘CTS’ signal enabled.
Software : transmission control by ‘XON’ and ‘XOFF’ enabled.

Capacity of receiving buffer

200 bytes

Escape sequence

for reception only

Functions

• Talker functions

Output of measurement data (ASCII, binary) and setting parameters (ASCII)

• Listener functions

Setting of measurement conditions, control of measurement start and stop (excluding the setting
and control of power on/off), specifying causes of ‘ESC S’ (output of a status byte)

Contents of ‘status’

Syntax error, chart end, completion of A/D conversion, operations of internal timer, end of data
saving/reading, incomplete measurement during computation