14.1 Specifications of DR130/DR231/DR241 (Style S3)

Standard Computation Functions

Kinds of computation

Difference between channels, linear scaling (scaling) and moving average.

Scaling

Scalable range: DC voltage, TC, RTD, contact
Scaling range: –30000 to +30000
Measurement accuracy for scaling: measurement accuracy for scaling (digits) = measurement
accuracy (digits)

× scaling span (digits) / measurement span (digits) + 2 digits. Numbers below

the decimal point are discarded.
(Example)
Measurement accuracy when the following settings are made:
Measuring range: 6 VDC, Measurement span: 1.000 to 5.000 V, Scaling span: 0.000 to 2.000,
Measured value: 5 V
+/-(0.05

Ч 5 + 2) Ч 2000/4000 + 2 = +/-3.125

Measurement accuracy = +/-4 digits = +/-0.004 V

Moving average

Moving average results for between 2 to 64 scans are computed.

Data Save/Load Function

Media for data save/load

Buffer memory (internal SRAM)

Capacity: 512 KB
Data backup: Around 10 years (backup with lithium battery, at room temperature while power
is off)
Specified data length: 10, 20, 30, 40, 50, 100, 200, 300, 400, 500, 1k, 2k, 3k, 4k, 5k, 10k, 20k,
30k, 40k, and 50k data/ch (Total memory lengt must be within the free memory size.)

3.5-inch floppy disk

Number of drives: 1
Disk types: 2HD, 2DD
Supported formats: 1.2 MB, 1.44 MB, and 720 KB

Applicable data

Setting values, measured values, and computed values (only possible whe optional math function
is specified)

Method to save to the floppy disk

Copies data stored in the buffer memory to the floppy disk, except for setting values which can be
directly saved to the floppy disk.

Method to load from floppy disk

Copies data from the floppy disk to the buffer memory except for settin values which can be
directly loaded from the floppy disk.

Printing and outputing loaded data

Able to print captured data saved in the buffer memory or output to a communication interface.

Data save format

Setting values: ASCII
Measured/computed values: binary (except ASCII (CSV Format) is also possible when saving to
floppy disk.)

Data capacity

Setting values

DR130 : Maximum about 42 KB (in case when saving the setting values of an operation mode

with 30ch inputs and 30ch computations.)

DR231/DR241 : Maximum about 50 KB (in case when saving the setting values of an

operation mode with 30ch inputs and 30ch computations.)

Measured values

Binary data: 2 bytes / 1 data
ASCII data: 12bytes / 1 data

Computed values

Binary data: 4 bytes / 1 data
ASCII data: 12 bytes / 1 data

Equation to calculate the total data capacity

256 + 64

× (number of measured ch + number of computation ch) + (number of measured ch ×

2 + number of computation ch

Ч 4 + 6) Ч specified data length