Sonic Alert Digital Program Controller DCP301 User Manual

Page 157

Advertising
background image

Chapter 10. SPECIFICATIONS

10-2

Over-range detection

threshold

Cold junction compensa-

tion accuracy

Influence of surrounding

temperature on cold

junction compensation

Cold junction compensa-

tion system

Scaling

Square root extraction

Linearization table approximation

Input bias

Digital filter

Number of inputs

Types of connectable

outputs

Terminal voltage (open)

Terminal current (short-

circuit)

Allowable contact resistance

(dry contact)

Residual current (allowable

open-collector ON)

Leakage current (allowable

open-collector OFF)

Parallel connection to

other instruments

Assignments (fixed)

Assignments (variable)

Input sampling cycle

ON detection min. hold time

Upper display

Lower display

Program No. display

Segment No. display

Profile display

Status displays

Operation keys

Loader connector port

Input

External

Switch

(RSW)

Input

Indication

/Program-

mer

110%FS min.: Upscaled

-10%FS max.: Downscaled (Note that F50 range is not downscaled. Lower readout

limit of B18 range is 20

°

C, 68

°

F.)

±

0.5

°

C (under standard conditions)

±

0.2

°

C (at 0 to 50

°

C range)

Internal/external (0

°

C only) compensation selectable

-1999 to 9999U (settable at dc voltage and dc current ranges. Reverse scaling and

decimal point repositioning possible)

Dropout 0.1 to 10.0%, Possible by dc current and voltage ranges

12 (both line ends fixed, 11 points variable)

-1000 to 1000U variable

0.0 to 120.0s variable (filter OFF at 0.0)

12

Dry contacts (relay contact) and open-collector (current sink to ground)

10.4 to 12.6V (under operating conditions) across common terminal (terminal

25

) and

each input terminal

5.0 to 6.6mA across each terminal (under operating conditions)

ON:

700

max. (under operating conditions)

OFF:

10k

min. (under operating conditions)

3V max. (under operating conditions)

0.1mA max. (under operating conditions)

Can be connected to DCP301/302 series.

RUN, HOLD, RESET, ADV, program No.

FAST, PV start, AT, AUTO/MANUAL, G.Soak cancel, reverse/direct action

0.1s

0.2s (program No. 0.4s)

Green 4-digit, 7-segment LED

This normally displays PV values. Item codes are displayed in parameter setup.

Orange 4-digit, 7-segment LED

This normally displays SP values. Setting values are displayed in parameter setup.

Green 2-digit, 7-segment LED

This normally displays program No.

Green 2-digit, 7-segment LED

This normally displays segment No. Item Nos. are displayed in parameter setup, and

alarm No. is displayed when alarm occurs.

6 orange LEDs

Displays program pattern rise, soak and fall tendencies.

22 round LEDs

Modes:

RUN, HLD, MAN, PRG (green)

Display details: PV, SP, OUT, TM, CYC (green)

Battery voltage: BAT (red) (blinks at low voltage)

Status:

AT, OT1, OT2, OT3 (orange)

Events:

EV1, EV2, EV3, T1, T2, T3, T4, T5 (orange)

13 rubber keys

1 (dedicated cable with stereo miniplugs)

Item

Specification

Advertising