2 setting the filter (pen model) – Omega RD100B User Manual

Page 110

Advertising
background image

6-2

6.2

Setting the Filter (Pen Model)

This section explains the details of setting a filter on the measurement channels on the

pen model.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or SHIFT +

key to select Aux and then press the

key.

3.

Press the

key or SHIFT +

key to select Filter and then press the

key.

4.

Set each item and press the

key.

Use the

key or SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up

to then are cancelled, and the display returns to a higher level menu.

5.

When the Setting complete screen appears, do either of the following:

Press the

key to set other channels.

If you are done, press the ESC key.

6.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Set the channel range.

Set the filter time constant.

The new setting takes effect.

ESC/?

First channel

Last channel

Set=Aux

Aux=Filter

CH=01-01

Resp.Time=Off

01-01 Channel
Setting complete

Explanation

Filter Time Constant

The time constant is selectable from 2 s, 5 s, and 10 s. Select Off when not using the

filter.

<Related Topics>

Enabling the filter function: Section 7.12