12-19

Specifications

12

Standard Performance

Item

Specifications

Measurement and recording accuracy

The following specifications apply to operation of the recorder under standard operation conditions:
Temperature:

23 ± 2°C

Humidity:

55% ± 10%RH

Power supply voltage:

90 to 132 or 180 to 264 VAC

Power supply frequency: 50/60 Hz ± 1%
Warm-up time:

At least 30 minutes.

Other ambient conditions such as vibration should not adversely affect the operation

Measurement (Digital Display)

Input Type

Recording (Trend)

Highest Res.

Measurement Accuracy

Recording Accuracy

Resolution

10 µV

Measurement
accuracy
± (0.3% of
recording span)

1 mV

±(0.15% of rdg + 1)

±(0.1% of rdg + 2 digits)

except R,S: 0 to 100°C, ±3.7°C

100 to 300°C, ±1.5°C

B: 400 to 600°C, ±2°C

except ±(0.15% of rdg + 1°C)

0.1°C

1 mV

10 mV

Accuracy not warranted for
values less than 400°C

Range Type

±(0.1% of rdg + 2 digits)

±(0.1% of rdg + 2 digits)

10 mV
1 mV

±(0.15% of rdg + 0.7°C)

for –200 to –100°C

±(0.15% of rdg + 0.5°C)

±(0.15% of rdg + 0.5°C)
except ±(0.15% of rdg + 0.7°C)
for J: –200 to –100°C

±(0.15% of rdg + 0.7°C)

±(0.15% of rdg + 1°C)

±(0.15% of rdg + 0.5°C)
except ±(0.15% of rdg + 0.7°C)
for L: –200 to –100°C

Pen model:
Dead band 0.2%

of recording span

Dot model:

Resolution 0.1 mm

10 µV
100 µV

±(0.2% of rdg + 1°C)

±(0.15% of rdg + 0.3°C)

DC voltage

TC

DC voltage

(excludes
RJC
accuracy)

RTD

ON/OFF input

20 mV

60 mV

200 mV

2 V

6 V

20 V

R

1-5 V

S

B

K

E

J

N

L

W

T

U

50 V

WRe

Pt100
JPt100
Level
Contact

Threshold level (2.4 V)
accuracy ± 0.1 V
≤ 1 KΩ: ON, ≥ 100 kΩ: OFF
(Parallel capacitance
0.01 µF or less)

Measuring accuracy in case of linear scaling

Accuracy during scaling (digits) = measuring accuracy (digits) × multiplier + 2 digits (rounded up)
where the multiplier = scaling span (digits)/measuring span (digits).
Example: Assuming that
• range type:

1-5 V

• measuring span:

1.000 to 5.000 V

• scaling span:

0.000 to 2.000 Then,

Measuring accuracy = ±(0.1% × 5 V + 2 digits)

= ±(0.005 V [5 digits] + 2 digits)
= ±7 digits

Multiplier = {2000 digits (0.000 to 2.000)}/4000 digits (1.000 to5.000) = 0.5
Accuracy during scaling = ±(7 × 0.5 + 2) digits = 6 digits (rounded up)

Reference junction compensation

Internal/External selectable for each channel

12.6 General Specifications