Omega RD100B User Manual

Page 112

Advertising
background image

6-4

6.4

Setting Recording Zones for Each Channel
(Zone Recording)

This section explains the details of setting the recording zone for each measurement

channel.

Procedure

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Press the

key or SHIFT +

key to select Aux and then press the

key.

3.

Press the

key or SHIFT +

key to select Zone and then press the

key.

4.

Set each item and press the

key.

Use the

key or SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up

to then are cancelled, and the display returns to a higher level menu.

5.

When the Setting complete screen appears, do either of the following:

Press the

key to set other channels.

If you are done, press the ESC key.

6.

Hold down the

MENU

key for 3 seconds to return to Operation mode.

Set the channel range.

Set the leftmost value of the
recording zone.

Set the rightmost value of the
recording zone.

The new setting takes effect.

First channel Last channel

ESC/?

Set=Aux

Aux=Zone

CH=01-01

Left= 0

Right=100

01-01 Channel
Setting complete

Explanation

Zone

The recording zone can be set in the following range.

Leftmost position:

0 mm to 95 mm

Rightmost position: 5 mm to 100 mm

Set a larger value for the rightmost position than the leftmost position.

Set at least 5 mm for the zone width.

Note

If the zone is set to a value less than 40 mm, scale printout of periodic printout cannot be

executed.