1-18

Alarm Printout

Alarm information is printed when an alarm occurs or releases.

Time of alarm occurrence/release

Indicates that there are alarms that are not
printed because the alarm printout buffer is full.

Level number

Alarm type

Channel No. or tag

: Alarm occurrence, : Alarm release

• The print condition can be set to (1) print when alarms occur and release, (2) print

only when alarms occur, or (3) do not print.

• Alarms that occur while an alarm printout is in progress are temporarily saved to the

buffer memory in a printout-wait condition. Alarms are cleared from the buffer

memory when they are printed.

• The number alarms that can be stored in the buffer is 8 and 12 on the pen model and

dot model, respectively. Alarms that occur while the buffer is full are not printed. A

buffer overflow mark is printed when there are alarms that cannot be printed because

the buffer is full.

• The time printout format can be selected.

Hour:Minute

Type

* The format of year, month, and day varies depending on the setting (see the next page).

Printout format

Type

Printout format

Hour:Minute:Second

Month:Day:
Hour:Minute

Year:Month:Day:
Hour:Minute:Second

Month:Day:
Hour:Minute:Second

10 : 00
Nov. 09
10 : 00
Nov. 09. 2004
10 : 00 : 00

10 : 00 : 00
Nov. 09
00 : 00 : 00

Periodic Printout

Measured values and other items are printed at the preset interval.

• Printout Contents (for details, see appendix 1)

• Date/time, time ticks (marks that indicate the positions of the date/time on the chart

paper); measured values, alarm status, scale (leftmost and rightmost values of

span) and recording color (pen model) for each channel; and chart speed can be

printed. When pen offset is being executed on the pen model, the pen offset marks

are printed. On the pen model, if a time tick is not printed at the correct position, a

time tick cancel mark (!) is printed.

• Printout of measured values and alarm status can be enabled or disabled for each

channel.

• Printout of the scale and recording colors (pen model) can be enabled or disabled.

The scale can be printed when the recording zone is greater than or equal to 40 mm.

• The measured values for each channel can be selected from the following data types.

Instantaneous value (measured value at the time of periodic printout)

Average value (average of the measured values over an interval)

Minimum value (minimum value of the measured values over an interval)

Maximum value (maximum value of the measured values over an interval)

Minimum value, maximum value, and average value

Sum value (summed value of the measured values over an interval)

• Interval

The printout interval can be set by specifying the value or set automatically in sync

with the chart speed. Periodic printout is disabled at the following chart speeds.

Pen model: Less than or equal to 9 mm/h or greater than or equal to 1600 mm/h

Dot model: Less than or equal to 9 mm/h or greater than or equal to 101 mm/h

1.4 Recording