9-14

Explanation

Timer No.

There are three timers.

Periodic: The timer for determining the periodic printout interval. The computed result

is printed every interval, and the computed TLOG value is reset.

1:

The timer of timer No. 1. Specify whether the computed value is printed every

interval and whether to reset the computed value.

2:

The timer of timer No. 2. Specify whether the computed value is printed every

interval and whether to reset the computed value.

• Timer operation

The table below shows how each timer operates.

Item

1 and 2*

Application
Target channel

Timer mode**
Reset computed TLOG value
Print when computation
stops

TLOG computation
Computation channels

Absolute or relative time mode
Yes/No selectable
Yes

Timer #

Periodic

Periodic
Measurement and
computation channels
Absolute time mode
Yes
No

*

For the timer 1 and 2 settings, see section 9.12.

** Absolute time mode: Keeps time from the reference time.

Relative time mode: Starts the time in sync with the starting of the computation.

SUM scale

Set the sum scale when using statistical calculation (TLOG.SUM).

For the definition of the term “sum scale,” see page 7-14.

Off:

Simply sums the computed values.

/s:

Sums by converting the computed values to a value over 1 second.

/min: Sums by converting the computed values to a value over 1 minute.

/h:

Sums by converting the computed values to a value over 1 hour.

/day: Sums by converting the computed values to a value over 1 day.

<Related Topics>

Setting the periodic printout timer: Section 7.8

Setting Timer 1 and Timer 2: Section 9.12

9.6 Specifying the Timer Used in Statistical Calculations (TLOG)