12-16

Item

Specifications

Measurement and recording accuracy

Input Type

Measurement Accuracy

Recording Accuracy

PR40-20

0 to 450

°C Not

warranted

450

to

750

°C

±(0.9% of rdg + 3.2°C)

750

to

1100

°C ±(0.9% of rdg + 1.3°C)

1100 to 1900

°C ±(0.9% of rdg + 0.4°C)

PLATINEL

±(0.25% of rdg + 2.3°C)

NiNiMo

±(0.25% of rdg + 0.7°C)

W/WRe26

0 to 400

°C

±15.0°C or less

400 to 2400

°C ±(0.2% of rdg + 2.0°C)

Type N(AWG14)

±(0.2% of rdg + 1.3°C) Measurement

Kp vs Au7Fe

0 to 20 K

±4.5 K

accuracy

20 to 300 K

±2.5 K

±(0.3% of

Pt25

±(0.15% of rdg + 0.6°C) recording

span)

Pt50

±(0.3% of rdg + 0.6°C)

Ni100(SAMA)

Ni100(DIN)

±(0.15% of rdg + 0.4°C)

Ni120

J263*B

0 to 40 K

±3.0 K

40 to 300 K

±1.0 K

Cu53

±(0.15% of rdg + 0.8°C)

Cu100

±(0.2% of rdg + 1.0°C)

Note) No RJC for PR40-20 (fixed to 0

°C)

Remote Control 5 Points (/R1)

Item

Specifications

Number of input terminals

5

Input type

Isolated from the main circuitry through a photocoupler, built-in isolated power supply for the input
terminals, and shared common

Input type and signal level

Voltage-free contact

Contact closed at 200 Ω or less and contact open at 100 kΩ or greater

Open collector

ON voltage: 0.5 V or less (sink current 30 mA or more), leakage current when OFF: 0.25 A or less

Signal type

Edge, level, or trigger (250 ms or more)

Operation

Perform the following control operations by applying a given signal to the remote signal input
terminal.
• Start/stop recording (edge)
• Alarm ACK (trigger)
• Set the clock (adjust the clock to the nearest hour with the contact input, trigger)

Time When Signal Is Input

Procedure

00 min 00 s to 01 min 59 s

Truncates the minutes and seconds. Example: 10:00:50 → 10:00:00

58 min 00 s to 59 min 59 s

Rounds up the minutes and seconds. Example: 10:59:50 → 11:00:00

02 min 00 s to 57 min to 59 s No operation

• Manual printout (trigger)
• Message 1 to 5 printout (trigger)
• Switch the chart speed (level)
• Start/stop computation (/M1 option) (edge)
• Reset computation (/M1 option) (trigger)

12.5 Specifications of Optional Functions