5-3

Frequently Used Setup Operations (Setting Mode)

5

Linear Scaling

Select Scale.

Set the channel range.

Select Volt, TC, RTD, or DI.

Select the range type.

Set the left span value.

Set the right span value.

Set the left scale value.

Set the right scale value.

The new setting takes effect.

First channel Last channel

ESC/?

Set=Range

CH=01-01

Mode=Scale

Type=Volt

Range=2V

Span_L= -2.000

Span_R= 2.000

Scale_L= 0.00

Scale_R= 200.00

01-01 Channel
Setting complete

Description

• Scale Left and Scale Right

Selectable range (mantissa): –20000 to 30000

Decimal position: XXXXX, XXXX.X, XXX.XX, XX.XXX, X.XXXX

The decimal position is the position specified by scale left.

Note

The displayable/printable range of scale values is –19999 to 30000 excluding the decimal.

5.1 Setting the Input range